Volumetric-modulated Arc Therapy Lung Stereotactic Body Radiation Therapy Dosimetric Quality Assurance: A Comparison between Radiochromic Film and Chamber Array

机译:体积调制电弧疗法肺立体定向身体放射疗法剂量学质量保证:放射致变色膜与腔室阵列的比较

获取原文
           

摘要

Introduction: The aim of this work is to verify the use of radiochromic film in the quality assurance (QA) of volumetric-modulated arc therapy (VMAT) lung stereotactic body radiation therapy (SBRT) plans and compare the results with those obtained using an ion chamber array. Materials and Methods: QA was performed for 14 plans using a two-dimensional-array seven29 and EBT3 film. Dose values per session ranged between 7.5 Gy and 18 Gy. The multichannel method was used to obtain a dose map for film. Results: The results obtained were compared with treatment planning system calculated profiles through gamma analysis. Passing criteria were 3%/3 mm, 2%/2 mm and 3%/1.5 mm with maximum and local dose (LD) normalization. Mean gamma passing rate (GPR) (percentage of points presenting a gamma function value of 98.9% for all criteria when normalizing per maximum dose. When using LD, normalization was >92.7%. GPR values for the array were lower for all criteria; this difference being statistically significant when normalizing at LD, reaching 12% for the 3%/1.5 mm criterion. Conclusion: Both detectors provide satisfactory results for the QA of plans for VMAT lung SBRT. The film provided greater mean GPR values, afforded greater spatial resolution and was more efficient overall.
机译:简介:这项工作的目的是验证放射性变色膜在容积调制弧光治疗(VMAT)肺立体定向人体放射治疗(SBRT)计划的质量保证(QA)中的使用,并将结果与​​使用离子获得的结果进行比较室阵列。材料和方法:使用二维阵列seven29和EBT3膜对14个计划进行质量检查。每个疗程的剂量值介于7.5 Gy和18 Gy之间。多通道方法用于获得胶片的剂量图。结果:通过伽玛分析将获得的结果与治疗计划系统计算的轮廓进行比较。通过标准为3%/ 3 mm,2%/ 2 mm和3%/ 1.5 mm,最大和局部剂量(LD)标准化。平均伽玛通过率(GPR)(在每最大剂量归一化时所有标准的伽马函数值均为98.9%的点的百分比。当使用LD时,归一化> 92.7%。阵列的GPR值对于所有标准都较低;这在LD归一化时,差异具有统计学意义,对于3%/ 1.5 mm标准,达到12%结论:两种检测器均为VMAT肺SBRT计划的QA提供了令人满意的结果;该膜提供了更高的平均GPR值,提供了更高的空间分辨率并且整体上效率更高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号