Estimation of glycemic variability requires frequent measures of glucose and is greatly aided by continuous glucose monitoring (CGM); however, under real-world conditions, missing data or “gaps” of ≥ 10 minutes can occur in CGM data, affecting the reliability of certain estimates. Thus, we determined the magnitude of the gap problem as observed in a cohort of patients with type 2 diabetes and demonstrated an approach to fill the gaps. The approach takes the difference between readings before and after a gap and distributes the difference equally across the number of missing readings, as determined by the sensor's setting for reading frequency. The approach is easy to implement, conservative, and improves estimation of variability measures that reference time, namely, mean of daily differences and continuous overlapping net glycemic action.
机译:估计血糖变异性需要频繁测量葡萄糖,并且连续血糖监测(CGM)极大地帮助了血糖的变化。但是,在现实情况下,CGM数据中可能会出现数据丢失或“间隙”≥10分钟,从而影响某些估计的可靠性。因此,我们确定了在一组2型糖尿病患者中观察到的缺口问题的严重程度,并证明了一种填补缺口的方法。该方法采用间隙之前和之后的读数之间的差异,并将差异平均分配到缺失读数的数量上,这取决于传感器的读数频率设置。该方法易于实施,保守并且改进了以时间为参考的变异性度量的估计,即日差的平均值和连续重叠的净血糖作用。