首页> 外文期刊>Journal of Biomechanical Science and Engineering >Development of a Passive Knee Motion Simulator to Evaluate Deep Knee Flexion of Total Knee Prosthesis
【24h】

Development of a Passive Knee Motion Simulator to Evaluate Deep Knee Flexion of Total Knee Prosthesis

机译:被动膝关节运动模拟器的开发,用于评估全膝关节假体的深膝屈曲

获取原文
           

摘要

References(26) Cited-By(1) To evaluate motions and joint reaction forces of the total knee prosthesis during deep knee flexion such as kneeling and sitting straight, a simulator that reproduces the passive motion of the knee has been developed. The main feature of the simulator is that the knee joint is accurately and repeatedly moved from 0° to 180° flexion angle in six degrees of freedom. Moreover, the tibiofemoral and patellofemoral motion and joint reaction forces can be continuously measured. Assuming that muscular force is mainly generated with a quadriceps femoris muscle, the similar force vector is produced by two motors. The custom-designed posterior stabilized type total knee prosthesis which can be flexed to 180° has been inserted in the bone model and used for the evaluation. As a result, it was confirmed that this simulator was able to measure the motion and joint reaction force of the knee with high repetition accuracy. The experimental results showed the same tendency with those in the previously performed cadaveric experiment until 120° flexion. Hereafter, cadaver knee can be used for the evaluation of the knee prosthesis using the proposed simulator with high accuracy.
机译:参考文献(26)Cited-By(1)为了评估在深屈膝(如跪着和直坐)期间全膝关节假体的运动和关节反作用力,已开发出可再现膝关节被动运动的模拟器。该模拟器的主要特点是,膝关节可以在六个自由度内从0°到180°屈曲角度精确重复地移动。此外,可以连续测量胫股和pa股运动以及关节反作用力。假设肌肉力主要由股四头肌产生,则相似的力矢量由两个电机产生。可以弯曲到180°的定制设计的后稳定型全膝关节假体已插入骨骼模型中,并用于评估。结果,证实了该模拟器能够以高重复精度测量膝盖的运动和关节反作用力。实验结果显示出与先前进行的尸体实验相同的趋势,直到弯曲120°。此后,尸体膝盖可以使用所提出的模拟器高精度地用于膝盖假体的评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号