Erythrocyte Sedimentation Rate: Comparison Of Dispette ? System Using Sodium Citrate 3.8% Versus Physiological Saline Solution 0.85% As Diluent

机译:红细胞沉降率:Dispette比较柠檬酸钠3.8%对生理盐溶液0.85%作为稀释剂的体系

获取原文
           

摘要

The most widely used manual method for measuring erythrocyte sedimentation rate (ESR) is the Dispette ? System. The International Committee of Laboratory Standards establishes the dilution of the sample with sodium citrate 3.8% and offers the physiological saline solution 0.85% as an alternative. OBJECTIVE: To compare the system using two diluent solutions: sodium citrate 3.8% versus physiological saline (SSF) 0.85%. METHODOLOGY: We applied to 84 samples diluted with SSF and sodium citrate the procedure described by the commercial firm and the scheme of standardization and quality assurance used in the chair of Hematology, School of Bioanalysis. RESULTS: The mean for ESR with SSF and with sodium citrate were practically the same and the standard deviations were very similar. The bias or error is -0.04 mm / h. CONCLUSION: The values of ESR using sodium citrate 3.8% and SSF 0.85% are comparable.
机译:Dispette是最广泛使用的手动测量红细胞沉降率(ESR)的方法。系统。国际实验室标准委员会确定了用3.8%的柠檬酸钠稀释样品并提供0.85%的生理盐水溶液作为替代方法。目的:比较使用两种稀释液的系统:柠檬酸钠3.8%与生理盐水(SSF)0.85%。方法:我们将84份用SSF和柠檬酸钠稀释的样品应用于商业公司描述的程序,以及生物分析学院血液学教授所使用的标准化和质量保证方案。结果:SSF和柠檬酸钠的ESR平均值几乎相同,标准偏差非常相似。偏差或误差为-0.04 mm / h。结论:使用3.8%的柠檬酸钠和0.85%的SSF的ESR值具有可比性。