Absorbed dose during a CT exam can be estimated by the computed tomography dose index (CTDI). The aim of this study is to validate a simulated CTDI with measured data for a 16-slice CT system. To measure the CTDI we have used an ion chamber, electrometer and two polymethylmethacrylate (PMMA) phantoms (head and body). GATE (GEANT4 Application for Tomographic Emission) platform is a Monte Carlo code for medical physics applications, especially dosimetry studies. CTDI and CTDIwwere recorded to compare simulated results for different X-ray tube currents and voltages with experimental data at different positions in the phantom. The results obtained for all conditions applied show a difference less than 2.6% and 4.3% for head and body phantom, respectively. It was demonstrated that GATE is a good tool to simulate CTDIwvalues and to study of the influence of tube current and voltage on the absorbed dose.
机译:可以通过计算机断层摄影剂量指数(CTDI)估算CT检查期间的吸收剂量。这项研究的目的是用16层CT系统的测量数据验证模拟的CTDI。为了测量CTDI,我们使用了离子室,静电计和两个聚甲基丙烯酸甲酯(PMMA)人体模型(头部和身体)。 GATE(层析成像发射的GEANT4应用)平台是用于医学物理应用(尤其是剂量学研究)的蒙特卡洛代码。记录CTDI和CTDI,以将不同X射线管电流和电压的仿真结果与模型中不同位置的实验数据进行比较。在所应用的所有条件下获得的结果表明,头部和体模的差异分别小于2.6%和4.3%。结果表明,GATE是模拟CTDIw值以及研究电子管电流和电压对吸收剂量的影响的良好工具。