A preliminary study of the external dose rates distribution for patients with prostate permanent 125I implants

机译:前列腺永久性125I植入物患者外部剂量率分布的初步研究

获取原文
           

摘要

The radiation dose rate distribution in the case of 137 patients after 125I prostatepermanent implants, performed at the Institute of Oncology, Bucharest, was experimentallydetermined. The study was performed on patients diagnosticated with prostate adenocarcinoma instages T1-T2. After the implant of radioactive sources the radiation level exposure were measured at2 meters distance from the anterior body surface. The value of the activities of the implanted sourcesranged from 580 to 1816 MBq while the exposure dose rates varied between 0.21 and 1.42 Sv/h.The radiation exposure levels suggest that patients are not presenting a radiation risk to the generalpublic and to the medical staff following 125I prostate permanent implants. The dose rates dependmainly on the implanted source activity, but also on the individual anatomical and physiologicalfactors and for this reason no direct correlation was found between the dose rates and the activities ofthe implanted sources.
机译:实验确定了在布加勒斯特肿瘤研究所进行的125 I前列腺永久植入后137例患者的辐射剂量率分布。该研究是针对诊断为T1-T2期前列腺腺癌的患者进行的。植入放射源后,在距人体前表面2米的距离处测量辐射水平。植入源的活动价值在580到1816 MBq之间,而暴露剂量率在0.21和1.42 Sv / h之间变化。辐射水平表明患者对普通公众和随后的医务人员没有辐射风险125I前列腺永久性植入物。剂量率主要取决于植入物源的活性,但也取决于各个解剖和生理因素,因此,在剂量率与植入物源的活性之间未发现直接相关性。