Abstract Venison colour was measured with a Minolta Chroma Meter CR200b portable filter colorimeter to give CIE (1976) L*a*b* (CIELAB) values the results were compared with those obtained with a Hunter LabScan 6000 scanning reflectance spectrocolorimeter and a sensory evaluation panel. The values obtained depended on the instrument used, but the variations from one instrument to another were systematic, enabling the filter colorimeter to be calibrated against the reflectance spectrocolorimeter. Panel scores could be accurately predicted by a linear combination of CIELAB values (R 2(adj) = 0.86).
机译:摘要用Minolta色度计CR200b便携式滤光色度计测量鹿肉色,得出CIE(1976)L * a * b *(CIELAB)值,并将结果与​​使用Hunter LabScan 6000扫描反射光谱仪和感官评估板获得的值进行比较。所获得的值取决于所用仪器,但是从一种仪器到另一种仪器的变化是系统的,因此可以根据反射光谱仪对滤光片色度计进行校准。小组得分可以通过CIELAB值的线性组合来准确预测(R 2(adj)= 0.86)。