Comparative Analysis of the Equivital EQ02 Lifemonitor with Holter Ambulatory ECG Device for Continuous Measurement of ECG, Heart Rate, and Heart Rate Variability: A Validation Study for Precision and Accuracy

机译:用于动态测量心电图,心率和心率变异性的等值EQ02生命监测器与动态心电图动态心电图设备的比较分析:精度和准确性的验证研究

获取原文
           

摘要

Background: The Equivital (EQ02) is a multi-parameter telemetric device offering both real-time and/or retrospective, synchronized monitoring of ECG, HR, and HRV, respiration, activity, and temperature. Unlike the Holter, which is the gold standard for continuous ECG measurement, EQO2 continuously monitors ECG via electrodes interwoven in the textile of a wearable belt. Objective: To compare EQ02 with the Holter for continuous home measurement of ECG, heart rate (HR), and heart rate variability (HRV). Methods: Eighteen healthy participants wore, simultaneously for 24 h, the Holter and EQ02 monitors. Per participant, averaged HR, and HRV per 5 min from the two devices were compared using Pearson correlation, paired T -test, and Bland-Altman analyses. Accuracy and precision metrics included mean absolute relative difference (MARD). Results: Artifact content of EQ02 data varied widely between (range 1.93–56.45%) and within (range 0.75–9.61%) participants. Comparing the EQ02 to the Holter, the Pearson correlations were respectively 0.724, 0.955, and 0.997 for datasets containing all data and data with < 50 or < 20% artifacts respectively. For datasets containing respectively all data, data with < 50, or < 20% artifacts, bias estimated by Bland-Altman analysis was ?2.8, ?1.0, and ?0.8 beats per minute and 24 h MARD was 7.08, 3.01, and 1.5. After selecting a 3-h stretch of data containing 1.15% artifacts, Pearson correlation was 0.786 for HRV measured as standard deviation of NN intervals (SDNN). Conclusions: Although the EQ02 can accurately measure ECG and HRV, its accuracy and precision is highly dependent on artifact content. This is a limitation for clinical use in individual patients. However, the advantages of the EQ02 (ability to simultaneously monitor several physiologic parameters) may outweigh its disadvantages (higher artifact load) for research purposes and/ or for home monitoring in larger groups of study participants. Further studies can be aimed at minimizing the artifacts.
机译:背景:Equivital(EQ02)是一种多参数遥测设备,可提供实时和/或回顾性同步监视ECG,HR和HRV,呼吸,活动和温度。与动态心电图连续测量黄金标准的动态心电图不同,EQO2通过可穿戴皮带织物中交织的电极连续监测心电图。目的:比较EQ02和动态心电图(Holter)在家庭中连续测量心电图,心率(HR)和心率变异性(HRV)。方法:18名健康参与者同时佩戴Holter和EQ02监护仪24小时。使用Pearson相关性,配对T检验和Bland-Altman分析,比较了这两个设备的每位参与者每5分钟的平均HR和HRV。准确性和精度指标包括平均绝对相对差(MARD)。结果:EQ02数据的伪像含量在参与者(范围为1.93-56.45%)和参与者(范围为0.75-9.61%)之间差异很大。将EQ02与动态心电图进行比较,包含所有数据和分别具有<50或<20%伪像的数据的数据集的Pearson相关性分别为0.724、0.955和0.997。对于分别包含所有数据,伪影<50或<20%的数据集,通过Bland-Altman分析估计的偏差为每分钟?2.8,?1.0和?0.8节拍,而24小时MARD为7.08、3.01和1.5。在选择了包含1.15%伪像的3小时数据后,HRV的Pearson相关系数为0.786,以NN间隔(SDNN)的标准偏差进行测量。结论:尽管EQ02可以准确测量ECG和HRV,但其准确性和精确度高度依赖于伪影含量。这是个别患者临床使用的限制。但是,EQ02的优点(能够同时监视多个生理参数)可能超过其缺点(较高的伪影负荷)以用于研究目的和/或用于较大数量的研究参与者的家庭监护。进一步的研究可能旨在最大程度地减少伪影。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号