首页> 外文期刊>Fluids and Barriers of the CNS >Hydrodynamics of the Certas? programmable valve for the treatment of hydrocephalus
【24h】

Hydrodynamics of the Certas? programmable valve for the treatment of hydrocephalus

机译:Certas的流体力学?可编程阀治疗脑积水

获取原文
           

摘要

Background The new Certas? shunt for the treatment of hydrocephalus has seven standard pressure settings that according to the manufacturer range from 36 to 238 mmH2O, and an additional “Virtual Off” setting with an opening pressure >400 mmH2O. Information on actual pressure response and reliability of shunt performance is important in clinical application, especially the “Virtual Off” setting as a non-surgical replacement for shunt ligation. The objective of this study was to evaluate the in-vitro hydrodynamic performance of the Certas? shunt. Methods Six new Certas? shunts with proximal and distal catheters were tested with an automated, computerized test system that raised the pressure from zero to a maximum pressure and back to zero at each valve setting. Opening pressure and flow resistance were determined. Results For settings 1–7 the measured opening pressure range was 26 to 247 mmH2O, and the mean change in opening pressure for a one-step adjustment was between 33 and 38 mmH2O. For setting 8 (“Virtual Off”) the measured mean opening pressure was 494?±?34 mmH2O (range 451 to 556 mmH2O). The mean outflow resistance was 7.0?mmHg/ml/min (outflow conductance 17.9 μl/s/kPa). Conclusions The six shunts had similar characteristics and closely matched the manufacturer’s specifications for opening pressure at settings 1–7. The opening pressure for the “Virtual Off” setting was nearly 500 mmH2O, which is 100 mmH2O higher than the manufacturer’s specification of “>400” and should be functionally off for most patients with communicating hydrocephalus. Clinical studies are needed to evaluate if the CSF dynamic profile persists after implantation in patients.
机译:背景新的Certas?用于脑积水治疗的分流器有七个标准压力设置,根据制造商的不同,其范围为36至238 mmH2O,另外还有一个“虚拟关闭”设置,其开启压力> 400 mmH2O。有关实际压力响应和分流器性能可靠性的信息在临床应用中非常重要,尤其是将“虚拟关闭”设置作为分流器结扎的非手术替代品。这项研究的目的是评估Certas?的体外水动力性能。分流器。方法新增六个Certas?使用自动化的计算机测试系统对带有近端和远端导管的分流器进行测试,该系统会将压力从零升高到最大压力,然后在每个阀门设置时都恢复到零。确定了打开压力和流动阻力。结果对于设置1–7,测得的打开压力范围为26到247 mmH2O,并且一步调整的打开压力的平均变化在33到38 mmH2O之间。对于设置8(“虚拟关闭”),测得的平均打开压力为494?±?34 mmH2O(范围为451至556 mmH2O)。平均流出阻力为7.0?mmHg / ml / min(流出电导为17.9μl/ s / kPa)。结论六个分流器具有相似的特性,并且与制造商在设置1–7时打开压力的规格非常匹配。 “虚拟关闭”设置的打开压力接近500 mmH2O,比制造商的“> 400”规格高100 mmH2O,对于大多数患有沟通性脑积水的患者,应该在功能上关闭。需要进行临床研究以评估患者植入后CSF动态图是否持续。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号