Balloon rupture during stent deployment is rare, and can cause stent damage upon its retrieval. We described a case of inadequate drug-eluting stent expansion in a fibrotic and calcified mid right coronary artery lesion despite pre-treatment with high pressure balloon inflation, with rupture and entrapment of the balloon within the stent. Attempts of retrieval of the ruptured balloon caused severe deformity of the distal segment of the stent. The stent could not be crushed or retrieved, and was left in situ. Angiographic follow-up at 8 months time showed widely patent vessel.
机译:支架展开过程中的球囊破裂很少见,并且在其取回时可能导致支架损坏。我们描述了尽管用高压球囊扩张进行了预处理,但球囊破裂并在支架内截留的情况,但在纤维化和钙化的右冠状动脉中病变中药物洗脱支架扩张不足的情况。试图恢复破裂的球囊会导致支架远端部分严重变形。支架不能被压碎或收回,而是留在原地。 8个月时的血管造影随访显示广泛的专利血管。