首页> 外文期刊>Annals of Pediatric Cardiology >Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus
【24h】

Transcatheter interruption of large residual flow after device closure of “Type A” patent ductus arteriosus

机译:在关闭“ A型”动脉导管未闭的装置后,经导管中断大的残留流量

获取原文
           

摘要

We report a case of 3-year-old girl who had persistence of large residual flow following transcatheter closure of a 6 mm ‘Type A’ patent ductus arteriosus using a 12 × 10 mm duct occluder. Angiography revealed a large left-to-right shunt coursing through and exiting around the implanted device. Near total abolition of the residual shunt was achieved by initial implantation of an embolization coil within the duct occluder and subsequently an Amplatzer duct occluder (ADO II) adjacent to the duct occluder. This challenging case describes an additional technique of abolishing a large residual flow in and around a Nitinol duct occluder device.
机译:我们报道了一例3岁女孩,该女孩使用12×10 mm导管封堵器经导管闭合6 mm的“ A型”动脉导管未闭后,仍有大量残余血流持续存在。血管造影显示,有一个较大的从左到右的分流器在植入的装置中穿行并流出。通过在导管封堵器中先植入栓塞线圈,然后在导管封堵器附近先植入Amplatzer导管封堵器(ADO II),几乎完全消除了残留分流器。这个具有挑战性的案例描述了一种其他技术,该技术可消除镍钛诺管道封堵器装置内部及其周围的大量残留流量。