首页> 外文期刊>Analytical Sciences >Electrochemical Detector Using Carbon Fiber Working Electrode for Micro High Performance Liquid Chromatography
【24h】

Electrochemical Detector Using Carbon Fiber Working Electrode for Micro High Performance Liquid Chromatography

机译:碳纤维工作电极用于微型高效液相色谱的电化学检测器

获取原文
           

摘要

A tubular electrolytic cell of ultra micro volume, made of a single carbon fiber (7μm o.d., 15mm length) as the working electrode material and a fused silica tube (50μm i.d.) as the flow cell part, was designed for use as the electrochemical detector in micro high performance liquid chromatography (micro HPLC). The effective cell volume is about 0.03μl. This detector with carbon fiber was useful for detection of catecholamines and their metabolites. Their detection limits were about 1-2pg and the linear dynamic range was about 103. The detector was successfully applied to urinary catecholamine analysis by micro HPLC and to rapid detection of electroactive species in fast micro HPLC.
机译:设计了一种超微体积的管状电解池,它由单根碳纤维(外径为7μm,长度为15mm)作为工作电极材料,熔融石英管(内径为50μm)作为流通池部件而设计,用作电化学检测器在微型高效液相色谱(微型HPLC)中进行分析。有效细胞体积为约0.03μl。这种带有碳纤维的检测器可用于检测儿茶酚胺及其代谢物。它们的检出限约为1-2pg,线性动态范围约为103。该检测器已成功应用于微量HPLC分析尿儿茶酚胺和快速微量HPLC快速检测电活性物质。