Downward bias of conductivity based point-of-care hemoglobin measurement compared with optical methods

机译:与光学方法相比,基于电导率的即时点血红蛋白测量的向下偏差

获取原文
           

摘要

Background: Point-of-care (POC) arterial blood gas analysis (ABGA) is widely used for checking hemoglobin (Hb) level. However, there is the tendency of downward bias of conductivity-based POC ABGA Hb measurement compared with optical methods. Authors tried to correct that bias by linear regression equation. Methods: We retrospectively collected a total of 86 Hb result pairs during surgeries. Hb measured by the Sysmex XE-2100 in the laboratory was set as the gold standard and was compared with that measured by the GEM Premier 3500. Data were compared using the Bland-Altman analysis, the reliability of transfusion decision was assessed using three-zone error grid. The linear regression analysis was performed to find out the relation between the Hb results of POC ABGA and those of laboratory based test. Results: The bias of the Hb measured between Sysmex XE-2100 and GEM Premier 3500 was ?0.9 g/dl (P 0.001, 95% confidence interval, ?1.038 to ?0.665 g/dl). The percentage error was 16.4%. According to error grid methodology, zone A, B and C encompassed 89.5%, 10.5% and 0% of data pairs. After adjusting the POC ABGA Hb values, the bias of the Hb measured by two methods was 0 g/dl (P = 0.991). The percentage error was 18.2%. The zone A, B and C encompassed 91.9%, 8.1% and 0% of data pairs. Conclusions: Hb measurements obtained with reference to conductivity via a POC ABGA were significantly lower than those obtained via optical methods. This bias may deserve attention of anesthesiologists when POC ABGA Hb level is used as a transfusion guideline.
机译:背景:即时护理(POC)动脉血气分析(ABGA)被广泛用于检查血红蛋白(Hb)水平。但是,与光学方法相比,基于电导率的POC ABGA Hb测量存在下降的趋势。作者试图通过线性回归方程来校正该偏差。方法:我们回顾性收集了手术期间总共86对Hb结果。 Sysmex XE-2100在实验室中测得的血红蛋白被设置为金标准,并与GEM Premier 3500测得的Hb进行了比较。使用Bland-Altman分析对数据进行了比较,使用三区域评估了输血决策的可靠性错误网格。进行线性回归分析以找出POC ABGA的Hb结果与实验室测试的Hb结果之间的关系。结果:在Sysmex XE-2100和GEM Premier 3500之间测得的Hb偏差为0.9 g / dl(P <0.001,95%置信区间1.038至0.665 g / dl)。百分比误差为16.4%。根据误差网格方法,区域A,B和C涵盖了89.5%,10.5%和0%的数据对。调整POC ABGA Hb值后,通过两种方法测得的Hb偏差为0 g / dl(P = 0.991)。百分比误差为18.2%。区域A,B和C涵盖了91.9%,8.1%和0%的数据对。结论:通过POC ABGA参照电导率获得的Hb测量值明显低于通过光学方法获得的Hb测量值。当POC ABGA Hb水平用作输血指南时,这种偏倚可能值得麻醉师注意。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号