Accuracy of Implant Placement Utilizing Customized Patient Instrumentation in Total Knee Arthroplasty

机译:利用定制的患者器械在全膝关节置换术中植入植入物的准确性

获取原文
           

摘要

Customized patient instrumentation (CPI) combines preoperative planning with customized cutting jigs to position and align implants during total knee arthroplasty (TKA). We compared postoperative implant alignment of patients undergoing surgery with CPI to traditional TKA instrumentation for accuracy of implant placement. Twenty-five consecutive TKAs using CPI were analyzed. Preoperative CT scans of the lower extremities were segmented using a computer program. Limb alignment and mechanical axis were computed. Virtual implantation of computer-aided design models was done. Postoperative coronal and sagittal view radiographs were obtained. Using 3D image-matching software, relative positions of femoral and tibial implants were determined. Twenty-five TKAs implanted using traditional instrumentation were also analyzed. For CPI, difference in alignment from the preoperative plan was calculated. In the CPI group, the mean absolute difference between the planned and actual femoral placements was 0.67° in the coronal plane and 1.2° in the sagittal plane. For tibial alignment, the mean absolute difference was 0.9° in the coronal plane and 1.3° in the sagittal plane. For traditional instrumentation, difference from ideal placement for the femur was 1.5° in the coronal plane and 2.3° in the sagittal plane. For the tibia, the difference was 1.8° in the coronal plane. CPI achieved accurate implant positioning and was superior to traditional TKA instrumentation.
机译:定制的患者器械(CPI)将术前计划与定制的切割夹具结合在一起,以在全膝关节置换术(TKA)期间定位和对齐植入物。我们将接受CPI手术的患者的术后植入物对准与传统的TKA仪器进行了比较,以了解植入物放置的准确性。使用CPI分析了25个连续的TKA。术前用计算机程序对下肢进行CT扫描。计算肢体对准和机械轴。完成了计算机辅助设计模型的虚拟植入。术后获得冠状位和矢状位X线片。使用3D图像匹配软件,确定股骨和胫骨植入物的相对位置。还分析了使用传统仪器植入的25个TKA。对于CPI,计算了与术前计划的一致性差异。在CPI组中,计划和实际股骨放置之间的平均绝对差在冠状平面内为0.67°,在矢状面内为1.2°。对于胫骨对准,平均绝对差在冠状面为0.9°,在矢状面为1.3°。对于传统器械,与股骨理想放置位置的差异在冠状平面上为1.5°,而在矢状平面上为2.3°。对于胫骨,在冠状平面中的差异为1.8°。 CPI实现了精确的植入物定位,并且优于传统的TKA仪器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号