Evaluation of Scatter Fraction (SF) of a 16-Ring Positron Emission Tomography (PET) System

机译:16环正电子发射断层扫描(PET)系统的散射分数(SF)评估

获取原文
           

摘要

Scatter fraction (SF) characteristic of a newly constructed 16-ring PET system with a FOV of 20 cm was experimentally evaluated in 2D acquisition mode (ring difference ≤ 3) but without septa, using a 10 cm diameter phantom. The scatter fraction measured from the backprojected images and the sinogram data aiming to find the discrepancy between the two approaches and to test the camera performance operating without septa. The images were reconstructed by the filtered backprojection (FBP) technique using some different filter functions. SF measured from the reconstructed images agreed within uncertainties and had a mean value of 26.1±0.8% at 250-850 keV energy and 12 ns coincidence time window. Scatter fraction also measured from analysis of the PET sinogram was slightly lower, at 23.6 ±0.7%. In each case, the procedure used for estimating the background under the peak is only approximate, and this may account for the slight discrepancy between the two approaches. The results so far indicate that the system can be operated in 2D mode without septa. DOI: http://dx.doi.org/10.3329/bjmp.v4i1.14691 Bangladesh Journal of Medical Physics Vol.4 No.1 2011 81-86
机译:使用直径为10 cm的幻像,以二维采集模式(环差≤3)但没有隔片,以实验方式评估了FOV为20 cm的新型16环PET系统的散射分数(SF)特性。从反投影图像和正弦图数据中测得的散射分数旨在发现两种方法之间的差异,并测试无隔片的相机性能。使用一些不同的滤波函数,通过滤波反投影(FBP)技术重建了图像。从重建图像测得的SF在不确定性内一致,并且在250-850 keV能量和12 ns符合时间窗口下的平均值为26.1±0.8%。通过PET正弦图分析也测得的散射分数略低,为23.6±0.7%。在每种情况下,用于估计峰下背景的过程仅是近似的,这可能解释了两种方法之间的细微差异。到目前为止的结果表明,该系统可以在没有间隔的情况下以2D模式运行。 DOI:http://dx.doi.org/10.3329/bjmp.v4i1.14691孟加拉国医学物理学杂志2011年第4卷第1期81-86

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号