Influence of Blood on the Accuracy of Raypex 5 and Root ZX Electronic Foramen Locators: An In Vivo Study

机译:血液对Raypex 5和Root ZX电子孔定位器精度的影响:一项体内研究

获取原文
           

摘要

The aim of this study was to evaluate in vivo the accuracy of the Raypex 5 and Root ZX electronic foramen locators (EFLs) in the presence of blood in the root canal space. Forty single-canal teeth scheduled for extraction were selected. Access cavity was prepared and coronal enlargement was carried out. Approximately two drops of blood were collected by finger prick and injected into the root canal space. The electronic working length (EWL) of each tooth by each device was established twice before (NB group) and after (WB group) injecting blood into the root canal. The tooth was extracted and the actual working length (AWL) was determined. Data were analyzed using McNemar's test. The accuracy rates of Raypex 5 and Root ZX within 0.5 mm in the NB group were 88.9% and 91.5%, with 83.3% and 86.2% in the WB group, respectively. There were no significant differences between the accuracy of each EFL in the two groups (p>0.05). Considering the NB and WB groups, there were no statistically significant differences in the accuracy of the EFLs (p>0.05). The presence of blood in the root canal space did not influence the accuracy of the EFLs.
机译:这项研究的目的是在根管腔中存在血液的情况下评估Raypex 5和Root ZX电子孔定位器(EFL)的准确性。选择了计划拔除的40条单管牙齿。准备进入腔并进行冠状扩大。用手指点刺收集大约两滴血液,并将其注入根管间隙。将每个设备的每个牙齿的电子工作长度(EWL)确定在(NB组)和(WB组)向根管内注入血液之前和之后两次。拔出牙齿并确定实际工作长度(AWL)。使用McNemar的检验分析数据。在NB组中,Raypex 5和Root ZX在0.5 mm以内的准确率分别为88.9%和91.5%,在WB组中分别为83.3%和86.2%。两组每个EFL的准确性之间无显着差异(p> 0.05)。考虑到NB和WB组,EFL的准确性没有统计学上的显着差异(p> 0.05)。根管腔内血液的存在并不影响EFL的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号