Clinical evaluation of the sterilizing effect at different sampling times in the dental implants surgery using local air laminar flow system

机译:使用局部空气层流系统对种植牙手术中不同采样时间的灭菌效果进行临床评估

获取原文
           

摘要

Objective: The purpose of the study was to determine the germicidal effect in the dental implant surgery using local air laminar flow system at different sampling times. Methods: The air samples were collected at different time points: 10min after laminar flow system turned on, 10min, 45min after operation and 10min before operation finished. Plate exposure method was used to detect the bacteria number. The samples collected at the same time points without the local laminar flow system open were taken as the control.. Results:When the local air laminar flowsystemwas used, the bacterial concentration in the surgery was significantly lower (p<0.05). Meanwhile, the bacterial concentration in the laminar region was significantly lower than that among the peripheral region (p<0.05). The cleanliness at the time point of 45min after operation, in the laminar area reached the level of 10min after laminar flowsystemturned on.When compared with the laminar region, the bacterial concentration in peripheral areas were significantly increased both at time points of 10min after operation and 10min before operation finished (p<0.05).Conclusion: Local air laminar flow system can filter the air in the surgery, enabling the air cleanliness to reach ten thousand grade. There is close relationship between the cleaning time and the cleaning effect. To ensure the clean effect, the management of the staff and the materials should be strengthened.
机译:目的:本研究的目的是确定在不同采样时间使用局部空气层流系统在种植牙手术中的杀菌效果。方法:在不同时间点采集空气样本:层流系统开启后10min,手术后10min,手术后45min和手术结束前10min。用平板曝光法检测细菌数。以未打开局部层流系统的同一时间点采集的样本作为对照。结果:使用局部空气层流系统时,手术中细菌浓度明显降低(p <0.05)。同时,层状区域中的细菌浓度显着低于周围区域中的细菌浓度(p <0.05)。手术后45min时,层流系统开启后,层流区域的清洁度达到10min的水平。与层流区域相比,在手术后10min的时间点和周围区域的细菌浓度均显着增加手术结束前10min(p <0.05)。结论:局部空气层流系统可在手术中过滤空气,使空气清洁度达到万级。清洁时间和清洁效果之间有着密切的关系。为确保清洁效果,应加强人员和物资的管理。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号