首页> 外文期刊>IEEE Transactions on Biomedical Engineering >Fast detection of venous air embolism in Doppler heart sound using the wavelet transform
【24h】

Fast detection of venous air embolism in Doppler heart sound using the wavelet transform

机译:小波变换快速检测多普勒心音中的静脉空气栓塞

获取原文
获取原文并翻译 | 示例
           

摘要

The introduction of air bubbles into the systemic circulation can result in significant morbidity. Real-time monitoring of continuous heart sound in patients detected by precordial Doppler ultrasound is, thus, vital for early detection of venous air embolism (VAE) during surgery. In this study, the multiscale feature of wavelet transforms (WT's) is exploited to examine the embolic Doppler heart sound (DHS) during intravenous air injections in dogs. As both humans and dogs share similar physiological conditions, the authors' methods and results for dogs are expected to be applicable to humans. The WT of DHS at scale 2/sup j/(j=1,2) selectively magnified the power of embolic, but not the normal, heart sound. Statistically, the enhanced embolic power was found to be sensitive (P<0.01 at 0.01 ml of injected air) and correlated significantly (P<0.0005, /spl tau/=0.83) with the volume of injected air from 0.01 to 0.10 ml. A fast detection algorithm of O(N) complexity with unit complexity constant for VAE was developed (processing speed=8 ms per heartbeat), which confirmed the feasibility of real-time processing for both humans and dogs.
机译:将气泡引入全身循环会导致大量发病。因此,通过心前区多普勒超声检测到的患者中连续心音的实时监测对于手术期间早期发现静脉空气栓塞(VAE)至关重要。在这项研究中,小波变换(WT)的多尺度特征被用来检查狗静脉注射空气期间的栓塞多普勒心音(DHS)。由于人类和狗都具有相似的生理条件,因此作者的狗方法和结果有望适用于人类。 DHS的比例为2 / sup j /(j = 1,2)的WT有选择地放大了栓塞的力量,但没有放大正常的心音。从统计学上看,增强的栓塞能力是敏感的(在0.01 ml注入空气中P <0.01),并且与从0.01到0.10 ml的注入空气量显着相关(P <0.0005,/ spl tau / = 0.83)。开发了一种OAE复杂度快速检测算法,其中VAE具有单位复杂度常数(处理速度= 8 ms /心跳),这证实了对人和狗进行实时处理的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号