首页> 外文期刊>American Journal of Neuroradiology >Small-Vessel Stents for Intracranial Angioplasty: In Vitro Comparison of Different Stent Designs and Sizes by Using CT Angiography
【24h】

Small-Vessel Stents for Intracranial Angioplasty: In Vitro Comparison of Different Stent Designs and Sizes by Using CT Angiography

机译:小血管支架用于颅内血管成形术:使用CT血管造影术对不同支架设计和尺寸的体外比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

BACKGROUND AND PURPOSE: Our purpose was to evaluate whether CT angiography is a suitable alternative to conventional angiography in the evaluation of small-vessel stents for intracranial angioplasty. METHODS: CT angiographic appearances of 23 stents of different designs and sizes (2.0, 3.0, and 4.0 mm) were investigated after they were filled with a solution of 0.9% NaCl or diluted contrast medium. For each stent, artificial lumen narrowing (ALN) was measured, and the difference in the number of pixels with a Hounsfield value below 200 HU between the two filling states, DIFFHU<200, was calculated to provide an objective indicator of the size of the evaluable stent diameter. RESULTS: With a window width of 1500 HU at a window level of 400 HU, ALN ranged from 66.8% to 97.7% in the group of 2.0-mm stents and from 38.6% to 66.8% in the groups of 3.0- and 4.0-mm stents. For the 2.0-mm stents, DIFFHU<200 was zero. In the groups of 3.0- and 4.0-mm stents, DIFFHU<200 ranged from 0.3 to 6.7, corresponding to a diameter of 0.13–3.0 mm, when the pixel size was presupposed to be 0.449 mm. CONCLUSION: CT angiographic evaluation of small-vessel patency after stent placement is considerably impaired by ALN. Stent manufacturers should be aware of potential artifacts caused by their stents during noninvasive diagnostic studies such as CT angiography.
机译:背景与目的:我们的目的是评估在颅内血管成形术的小血管支架评估中, CT血管造影是否可以替代常规血管造影 。 sup>方法:在 填充0.9%的溶液后,研究了23个不同 设计和尺寸(2.0、3.0和4.0 mm)的支架的CT血管造影外观。 NaCl或稀释的对比度 中。对于每个支架,均测量了 的人造管腔变窄(ALN),并且两个填充状态之间的 Hounsfield值小于200 HU的像素数之差, DIFF HU <200 ,用于提供可评估支架直径大小的客观指标 结果:在 400 HU窗口水平下的1500 HU窗口宽度,在2.0 mm 支架组中,ALN范围从66.8%到97.7%,从38.6%到66.8不等在3.0毫米和4.0毫米 支架组中的百分比。对于2.0毫米支架,DIFF HU <200 为零。在3.0和4.0毫米支架的 组中,DIFF HU <200 的范围从 0.3到6.7,对应于直径0.13– 3.0 mm, ,当像素大小假定为0.449 mm时。 结论:支架置入后小血管通畅性的CT血管造影评价被ALN损害。 Stent 制造商应注意在进行 CT血管造影等非侵入性诊断研究过程中其支架引起的 潜在假象。

著录项

  • 来源
    《American Journal of Neuroradiology》 |2003年第8期|00001512-00001516|共5页
  • 作者单位

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

    From the Division of Neuroradiology, Department of Neurology, University of Heidelberg Medical Center, Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献