摘要

PurposeTomosynthesis is attracting attention as a low-dose tomography technology compared with X-ray CT. However, conventional tomosynthesis imaging devices are large and stationary. Furthermore, there is a limitation in the working range of the X-ray source during image acquisition. We have previously proposed the use of a portable X-ray device for tomosynthesis that can be used for ward rounds and emergency medicine. The weight of this device can be reduced by using a flat panel detector (FPD), and flexibility is realized by the free placement of the X-ray source and FPD. Tomosynthesis using a portable X-ray device requires calibration of the geometry between the X-ray source and detector at each image acquisition. We propose a method for geometry calibration and demonstrate tomosynthesis image reconstruction by this method.
机译:目的与X射线CT相比,断层合成作为一种低剂量断层扫描技术受到关注。然而,常规的断层合成成像装置较大且固定。此外,在图像获取期间X射线源的工作范围受到限制。先前我们已经提出使用便携式X射线设备进行断层合成,该设备可用于病房巡查和急诊医学。可以通过使用平板检测器(FPD)减轻设备的重量,并通过自由放置X射线源和FPD来实现灵活性。使用便携式X射线设备进行断层合成需要在每次图像采集时校准X射线源和检测器之间的几何形状。我们提出了一种几何定标方法,并演示了层析合成图像的重建方法。