首页> 美国卫生研究院文献>The Journal of Automatic Chemistry >International Federation of Clinical Chemistry (IFCC): Scientific DivisionCommittee on pH Blood Gases and Electrolytes: Guidelines for Transcutaneouspo2andpco2 Measurement
【2h】

International Federation of Clinical Chemistry (IFCC): Scientific DivisionCommittee on pH Blood Gases and Electrolytes: Guidelines for Transcutaneouspo2andpco2 Measurement

机译:国际临床化学联合会(IFCC):科学部pH血气和电解质委员会:经皮po2和pco2测量指南

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

This document provides guidelines for the terminology, methodology, and for the interpretation of data obtained from the use of skin (transcutaneous) po2 and pco2 electrodes. The transcutaneous technique has found special application in newborn infants. The causes of analytical bias with respect to arterial blood gas values, and imprecision obtained with transcutaneous pco2 electrodes, are reviewed. Electrode temperatures above 44°C should not be used routinely, and, at a measuring temperature of 44°C, the measuring site should be changed at least every 4 h to avoid skin burns.
机译:本文档提供了有关术语,方法论以及对使用皮肤(经皮)po2和pco2电极获得的数据的解释的指南。经皮技术已发现在新生儿中有特殊应用。回顾了关于动脉血气值分析偏倚的原因,以及经皮pco2电极获得的不精确性。电极温度不得超过44°C,并且在测量温度为44°C时,应至少每4小时更换一次测量位置,以免灼伤皮肤。