Static and Dynamic Differences in Fixation Stability between a Spacer Plate and a Small Stature Plate Fixator Used for High Tibial Osteotomies: A Biomechanical Bone Composite Study

机译:用于胫骨高位截骨术的垫片和小身板固定器之间固定稳定性的静态和动态差异:生物力学骨复合材料研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Background. The objective of the present study was to compare mechanical strength and stability of the newly designed spacer plate with the gold standard plate for the treatment of medial knee joint osteoarthritis. Materials and Methods. Ten fourth-generation tibial bone composites underwent a medial open-wedge high tibial osteotomy (HTO) according to standard techniques, using five TomoFix plates and five Contour Lock plates. Static compression load to failure and load-controlled cyclical fatigue failure tests were performed. Forces and horizontal displacements were measured; plastic deformations and dynamic stiffness were determined. Results and Discussion. In all samples, rotation of the tibial head and fracture of the opposite cortex were observed. Behaviors of the specimens under static loading were comparable between groups. Cyclic testing revealed lateral significant higher stiffness until failure for the Contour Lock compared to the TomoFix plate. No visible implant failure was observed in any group. Conclusion. Considering the static analysis, both plates offered sufficient stability under physiologic loads of up to 3000 N. The Contour Lock plate-fixated specimens showed a higher stability during the cyclic testing, supposedly due to the wider distance between the fixation screws.
机译:背景。本研究的目的是比较新设计的间隔板和金标准板的机械强度和稳定性,以治疗内侧膝关节骨关节炎。材料和方法。根据标准技术,使用五个TomoFix板和五个Contour Lock板,对十个第四代胫骨复合材料进行了内侧楔入式高位胫骨截骨术(HTO)。进行了静态压缩破坏荷载和受控循环疲劳破坏试验。测量力和水平位移。确定了塑性变形和动态刚度。结果与讨论。在所有样品中,均观察到胫骨头旋转和相对皮质的骨折。两组在静态载荷下的试样行为相当。循环测试显示,与TomoFix板相比,Contour Lock失效前的侧向刚度明显更高。在任何组中均未观察到可见的植入失败。结论。考虑到静态分析,两块板在高达3000 N的生理载荷下都具有足够的稳定性。Contour Lock板固定的标本在循环测试中显示出更高的稳定性,这可能是由于固定螺钉之间的距离更大。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号