首页> 美国卫生研究院文献>Physics and Imaging in Radiation Oncology >In vivo dosimetry with an electronic portal imaging device for prostate cancer radiotherapy with an endorectal balloon
【2h】

In vivo dosimetry with an electronic portal imaging device for prostate cancer radiotherapy with an endorectal balloon

机译:在体内剂量测定中具有用于前列腺癌放射疗法的电子门静脉成像装置

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Electronic portal imaging device-based in vivo dosimetry with a commercial product was performed for 10 prostate cancer patients treated with an air-filled endorectal balloon. With the conventional in vivo method the verification results were outside of our clinical acceptance criteria for all patients. The in aqua vivo method, originally developed for lung cancer treatments, proved to be a practical solution to this problem. On average the percentage of points within γ agreement of 3% and 3 mm significantly improved from 90.9% ± 2.5% (1SD) for the conventional in vivo method to 99.0% ± 1.0% (1SD) for the in aqua vivo method.
机译:基于电子门户成像装置的基于体内剂量测定的具有商业产品的10个前列腺癌患者进行,所述前列腺癌患者用充满空气的山羊处理。随着常规体内方法,验证结果超出了所有患者的临床验收标准之外。 AQUA体内方法最初为肺癌治疗方法开发,证明是对此问题的实用解决方案。平均γ协议内的点数为3%和3mm的百分比从90.9%±2.5%(1SD)的常规方法显着提高到AQHA体内方法的99.0%±1.0%(1SD)。