A comparative assessment of the accuracy of electronic apex locator (Root ZX) in the presence of commonly used irrigating solutions

机译:电子顶点定位器精度的比较评估(根ZX)存在常用的灌溉解决方案

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objectives: This study aimed to evaluate the accuracy of Root ZX in determining working length in presence of normal saline, 0.2% chlorhexidine and 2.5% of sodium hypochlorite. Material and Methods: Sixty extracted, single rooted, single canal human teeth were used. Teeth were decoronated at CEJ and actual canal length determined. Then working length measurements were obtained with Root ZX in presence of normal saline 0.9%, 0.2% chlorhexidine and 2.5% NaOCl. The working length obtained with Root ZX were compared with actual canal length and subjected to statistical analysis. Results: No statistical significant difference was found between actual canal length and Root ZX measurements in presence of normal saline and 0.2% chlorhexidine. Highly statistical difference was found between actual canal length and Root ZX measurements in presence of 2.5% of NaOCl, however all the measurements were within the clinically acceptable range of ±0.5mm. Conclusion: The accuracy of EL measurement of Root ZX within±0.5 mm of AL was consistently high in the presence of 0.2% chlorhexidine, normal saline and 2.5% sodium hypochlorite. Clinical significance: This study signifies the efficacy of ROOT ZX (Third generation apex locator) as a dependable aid in endodontic working length. > Key words:Electronic apex locator, working length, root ZX accuracy, intracanal irrigating solutions.
机译:目的:本研究旨在评估Root ZX在生理盐水,0.2%洗必泰和2.5%次氯酸钠存在下确定工作长度的准确性。材料和方法:使用六十颗拔出的,单根,单管的人类牙齿。在CEJ处对牙齿进行了装饰,并确定了实际的运河长度。然后,在0.9%的生理盐水,0.2%的洗必太和2.5%的NaOCl存在下,使用Root ZX获得工作长度测量值。将Root ZX获得的工作长度与实际根管长度进行比较,并进行统计分析。结果:在生理盐水和0.2%洗必泰存在的情况下,实际根管长度与根ZX测量值之间无统计学差异。在存在2.5%NaOCl的情况下,实际根管长度与根ZX测量值之间存在高度统计学差异,但是所有测量值均在±0.5mm的临床可接受范围内。结论:在存在0.2%的洗必泰,生理盐水和2.5%的次氯酸钠的情况下,AL的±0.5 mm以内的Root ZX的EL测量准确度始终很高。临床意义:这项研究表明ROOT ZX(第三代根尖定位器)作为根管治疗长度的可靠辅助工具的功效。 >关键字:电子根尖定位器,工作长度,根ZX精度,管内冲洗液。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号