The Effects of Sterilization Procedures on the Cutting Efficiency of Endodontic Instruments: A Systematic Review and Network Meta-Analysis

机译:灭菌程序对牙髓仪器切割效率的影响:系统审查与网络荟萃分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Sterilization processes guarantee the sterility of dental instruments but can negatively affect instrument features by altering their physical and mechanical properties. The endodontic instrumentation can undergo a series of alterations, ranging from corrosion to variation in the cutting angle and then changes in the torsional properties and torsional fatigue resistance. This systematic literature review and meta-analysis aims to investigate alterations to the cutting efficiency of endodontic instruments that are induced by procedures for their disinfection and sterilization. Methodologies adopted for this investigation follow the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analysis) guidelines. The following were used as search terms on PubMed and Scopus: “endodontic sterilization”, “endodontic autoclave”, “cyclic fatigue”, “torsional”, “cutting efficiency”, “sterilization”, “surface characteristics”, and “corrosion”. At the end of the selection process, 36 articles were identified, and seven of them are included in this systematic review. The results of a meta-analysis conducted for the use of 10 autoclaving cycles shows a standardized mean difference (SMD) of 0.80 with a p-value equal to 0.04 with respect to effect on cutting efficiency. The network meta-analysis, through direct and indirect comparison between the different autoclave cycles (0, 1, 5, 10, and 15 cycles), revealed that treatment involving 15 autoclave cycles produced the most robust results in terms of having the greatest effects in terms of altered cutting efficiency with a probability of 57.7% and a SUCRA (surface under the cumulative ranking) of 80%. The alterations in the effects on cutting efficiency appear to be triggered after five cycles of sterilization by heat (autoclave). In conclusion, the meta-analysis of the data indicates that the autoclave sterilization protocol must not be repeated more than five times to preserve cutting efficiency. Within the limitations of this review, we can therefore establish that sterilization by autoclaving alone results in steel and NiTi instruments becoming less efficient in cutting after five cycles, as measured by a reduction in cutting efficiency.
机译:灭菌过程能够通过改变其物理和机械性能来保证牙科仪器的无菌性,但可以对仪器特征产生负面影响。牙髓仪器可以进行一系列改变,从腐蚀到切割角度的变化,然后改变扭转性能和扭转疲劳性。该系统文献综述和META分析旨在调查通过程序消毒和灭菌的程序诱导的牙髓仪器的切削效率的改变。本调查采用的方法遵循PRISMA(首选报告项目进行系统评价和荟萃分析)指导方针。以下用作PubMed和Scopus上的搜索条件:“牙髓灭菌”,“胸腺式高压釜”,“循环疲劳”,“扭转”,“切割效率”,“灭菌”,“表面特征”和“腐蚀”。在选择过程结束时,鉴定了36篇文章,其中七种文章包括在该系统审查中。用于使用10个高压灭菌循环的Meta分析的结果显示出0.80的标准化平均差(SMD),P值相对于切割效率的影响等于0.04。通过不同的高压釜循环(0,1,5,10和15次循环之间的直接和间接比较的网络元分析显示,涉及15个高压釜循环的治疗在具有最大的效果方面产生了最强大的结果改变的切削效率的概率为57.7%,SUCRA(累积排名下的表面)为80%。在热量(高压釜)的五个杀菌循环后,似乎在切割效率的影响中的改变似乎被触发。总之,数据的荟萃分析表明,高压灭菌灭菌方案不得重复超过五次以保持切削效率。在本综述的局限内,我们可以通过单独通过降低切割效率的降低来确定通过钢和NITI仪器的钢和NITI仪器的高压灭菌来确定灭菌。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号