首页> 美国卫生研究院文献>Indian Pacing and Electrophysiology Journal >Late presentation of recurrent syncope after permanent pacemaker implantation due to Lead−Header malapposition
【2h】

Late presentation of recurrent syncope after permanent pacemaker implantation due to Lead−Header malapposition

机译:永久性起搏器植入后的经常性晕厥的后期呈现因铅球发作原因

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Permanent pacemaker (PPM) malfunction due to electrical connection problems such as a loose set screw or lead-header malapposition is extremely rare. We present a patient with complete heart block (CHB) who had PPM malfunction and recurrent syncope, late (14 months) after initial implantation, which was caused by the ventricular lead pin disengagement from the header resulting in oversensing due to noise, pacing inhibition and recurrent syncope. PPM due to lead-header malapposition this late after device implantation has previously not been reported.
机译:永久起搏器(PPM)由于电气连接问题,如松散的固定螺钉或引线头部MOLAPPION等电气连接问题非常罕见。我们呈现患有PPM故障和复发性晕厥的完整心脏块(CHB)的患者,初始植入后(14个月)是由心室引线脱离引起的,导致由于噪音,起搏抑制和衰退而导致监督经常性晕厥。 PPM由于铅头MARAPPONION之前,在设备植入后期未报告。