首页> 美国卫生研究院文献>Hip Pelvis >Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report
【2h】

Gross Trunnion Failure in the Bipolar Hemiarthroplasty; Raising Concern about Short Trunnion: A Case Report

机译:双极半轴成形术中的总耳搏失效;提高关于短暂耳轴的关注:案例报告

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

There have been some reports of gross trunnion failure (GTF) in total hip arthroplasty. Here, we report a case of GTF 19 years after bipolar hemiarthroplasty using a 28-mm head with a 14/16 taper bore. Compared to other GTF reports, the current case had some unusual aspects: bipolar hemiarthroplasty, 28-mm head, relatively late-onset, and no apparent findings of metallosis though a severe one was evident. A Computed tomography scout view provided valuable information in evaluating polyethylene, metal head, and neck inside the bipolar cup. The current report suggests a need for concerned regarding short trunnion length which may be associated with GTF.
机译:在总髋关节置换术中有一些关于总耳病失败(​​GTF)的报告。在这里,我们通过28 mm头用14/16锥形孔的头部报告双极半导形塑料术后GTF 19年的情况。与其他GTF报告相比,目前的案例有一些不寻常的方面:双极性半序成形术,28毫米,相对深刻的表现,并且虽然严重的是显而易见的金属异位的明显发现。计算机断层扫描侦察术视图提供了评估双极杯内的聚乙烯,金属头和颈部的有价值的信息。目前的报告表明,需要有关与GTF相关的短暂耳轴长度的关注。