Blend uniformity analysis using stream sampling and near infrared spectroscopy

机译:使用流采样和近红外光谱进行混合均匀度分析

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A near infrared spectroscopic method was developed to determine drug content in a 20% (wt/wt) ibuprofen and spray-dried hydous lactose blend. A blending profile was obtained after blending for 0.5, 1, 3, 5, 10, and 20 minutes. Stream sampling was used to collect about 20 blend samples at each of the blending times from a laboratory scale V-blender. The samples collected were used to develop a near infrared calibration model. The calibration model was then used to determine the drug content of unknown samples from 2 validation blends. The validation blends were not included in the calibration model; they were used to evaluate the effectiveness of the calibration model. A total of 45 samples from the 2 validation blends were predicted by the near infrared calibration model and then analyzed by a validated UV spectrophotometric method. The root mean square error of prediction for the first validation blend was 5.69 mg/g and 3.30 mg/g for the samples from the second blend. A paired t test at the 95% confidence level did not indicate any differences between the drug content predicted by the near infrared spectroscopy (NIRS) method and the validated UV method for the 2 blends. The results show that the NIRS method could be developed while the blending profile is generated and used to thoroughly characterize a new formulation during development by analyzing a large number of samples. The new formulation could be transferred to a manufacturing plant with an NIRS method to facilitate blend uniformity analysis.
机译:开发了一种近红外光谱法来测定20%(wt / wt)布洛芬和喷雾干燥的含水乳糖混合物中的药物含量。在共混0.5、1、3、5、10和20分钟后获得共混曲线。在实验室规模的V型混合器中,在每个混合时间,均使用流采样来收集约20个混合样品。收集的样品用于建立近红外校准模型。然后,使用校准模型从2种验证混合物中确定未知样品的药物含量。验证混合物未包含在校准模型中;他们被用来评估校准模型的有效性。通过近红外校准模型预测来自2种验证混合物的总共45个样品,然后通过经过验证的UV分光光度法进行分析。第一个验证混合物的预测均方根误差为5.69 mg / g,第二个混合物的样品的预测均方根误差为3.30 mg / g。在95%置信水平下的配对t检验未显示两种混合物的近红外光谱(NIRS)方法和经过验证的UV方法预测的药物含量之间的任何差异。结果表明,可以在生成混合曲线的同时开发NIRS方法,并可以通过分析大量样品在开发过程中彻底表征新配方。可以使用NIRS方法将新配方转移到生产工厂中,以促进混合均匀性分析。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号