Optimized saturation recovery protocols for T1-mapping in the heart: influence of sampling strategies on precision

机译:心脏中T1映射的优化饱和度恢复方案:采样策略对精度的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundT1-mapping has the potential to detect and quantify diffuse processes such as interstitial fibrosis. Detection of disease at an early stage by measurement of subtle changes requires a high degree of reproducibility. Initial implementation of saturation recovery (SR) T1-mapping employed 3-parameter fitting which was highly accurate but was quite sensitive to noise; 2-parameter fitting greatly reduced the sensitivity to noise at the expense of a small degree of systematic bias. A recently introduced implementation that uses a variable readout flip angle greatly reduces systematic errors in T1-measurement thereby making it feasible to use SR methods with 2-parameter fitting with improved accuracy and precision. SR T1 mapping techniques with multi-heartbeat recovery times have been proposed to better sample the T1 recovery curve, but have not been evaluated for 2-parameter fitting.
机译:BackgroundT1映射有潜力检测和量化弥散过程,如间质纤维化。通过测量细微变化在早期阶段检测疾病需要高度的可重复性。饱和恢复(SR)T1映射的最初实现使用3参数拟合,该拟合非常准确,但对噪声非常敏感。 2参数拟合以较小的系统偏差为代价,大大降低了对噪声的敏感性。最近引入的使用可变读出翻转角的实现方式极大地减少了T1测量中的系统误差,从而使使用具有2参数拟合的SR方法具有更高的精度和精度成为可能。已经提出了具有多心跳恢复时间的SR T1映射技术来更好地采样T1恢复曲线,但尚未进行2参数拟合评估。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号