Percutaneous transluminal angioplasty is a well-known treatment for arteriovenous fistula stenosis. Balloon rupture during endovascular procedures is a rare but possible complication. The bursting balloon itself does not cause a serious problem, but it can occasionally cause entrapment, especially in case of breakdown of the balloon catheter tip. Here, we present four cases of balloon rupture during angioplasty in the hemodialysis circuit. In three cases, the ruptured balloon catheter was removed by cutdown of access conduit, while in one case, tip of ruptured balloon catheter was migrated into the pulmonary artery and was removed surgically. The operator must attempt to reduce the risk of balloon rupture by gradually expanding the balloons under bursting pressure. If the balloon bursts, it should not be removed excessively and attempt should be made to remove it using endovascular techniques. Surgical removal is considered after careful evaluation of the condition of the balloon and vessel.
机译:经皮腔内血管成形术是动静脉瘘狭窄的一种众所周知的治疗方法。血管内手术中的球囊破裂是一种罕见但可能的并发症。破裂的球囊本身并不会引起严重的问题,但有时会引起夹带,特别是在球囊导管尖端破裂的情况下。在这里,我们介绍了血液透析回路中血管成形术中球囊破裂的四例。在三例中,破裂的球囊导管通过进入导管的切除而移除,而在一种情况下,破裂的球囊导管的尖端被移入肺动脉并通过手术移除。操作员必须设法通过在破裂压力下逐渐膨胀气球来降低气球破裂的风险。如果球囊破裂,则不应将其过度移除,而应尝试使用血管内技术将其移除。在仔细评估球囊和血管的状况后,考虑手术切除。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号