A Revised Point-to-Point Calibration Approach with Adaptive Errors Correction to Weaken Initial Sensitivity of Cuff-Less Blood Pressure Estimation

机译:一种修正的点对点校准方法该方法具有自适应误差校正功能可减弱无袖带血压估计的初始灵敏度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Initial calibration is a great challenge for cuff-less blood pressure (BP) measurement. The traditional one point-to-point (oPTP) calibration procedure only uses one sample/point to obtain unknown parameters of a specific model in a calm state. In fact, parameters such as pulse transit time (PTT) and BP still have slight fluctuations at rest for each subject. The conventional oPTP method had a strong sensitivity in the selection of initial value. Yet, the initial sensitivity of calibration has not been reported and investigated in cuff-less BP motoring. In this study, a mean point-to-point (mPTP) paring calibration method through averaging and balancing calm or peaceful states was proposed for the first time. Thus, based on mPTP, a factor point-to-point (fPTP) paring calibration method through introducing the penalty factor was further proposed to improve and optimize the performance of BP estimation. Using the oPTP, mPTP, and fPTP methods, a total of more than 100,000 heartbeat samples from 21 healthy subjects were tested and validated in the PTT-based BP monitoring technologies. The results showed that the mPTP and fPTP methods significantly improved the performance of estimating BP compared to the conventional oPTP method. Moreover, the mPTP and fPTP methods could be widely popularized and applied, especially the fPTP method, on estimating cuff-less diastolic blood pressure (DBP). To this extent, the fPTP method weakens the initial calibration sensitivity of cuff-less BP estimation and fills in the ambiguity for individualized calibration procedure.
机译:初始校准对于无袖带血压(BP)测量是一个巨大的挑战。传统的单点对点(oPTP)校准过程仅使用一个样本/点来获得处于平静状态的特定模型的未知参数。实际上,对于每个对象,诸如脉冲传播时间(PTT)和BP之类的参数在静止时仍然会有轻微的波动。传统的oPTP方法在选择初始值时具有很强的敏感性。然而,在无袖带BP驾驶中,校准的初始灵敏度尚未得到报道和研究。在这项研究中,首次提出了一种通过平均和平衡平静或和平状态的平均点对点(mPTP)配对校准方法。因此,在基于mPTP的基础上,通过引入惩罚因子,提出了一种因子点对点(fPTP)配对校准方法,以改善和优化BP估计的性能。使用oPTP,mPTP和fPTP方法,在基于PTT的BP监测技术中对来自21位健康受试者的总计超过100,000个心跳样本进行了测试和验证。结果表明,与传统的oPTP方法相比,mPTP和fPTP方法显着提高了BP估计性能。此外,mPTP和fPTP方法,尤其是fPTP方法,在估计无袖套舒张压(DBP)方面可以广泛推广和应用。在此程度上,fPTP方法削弱了无袖带BP估计的初始校准灵敏度,并填补了个性化校准程序的不确定性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号