首页> 美国卫生研究院文献>Journal of Clinical Microbiology >Ability of Two Commercially Available Assays (Abbott RealTime HIV-1 and Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Version 2.0) To Quantify Low HIV-1 RNA Levels (1000 Copies/Milliliter): Comparison with Clinical Samples and NIBSC Working Reagent for Nucleic Acid Testing Assays
【2h】

Ability of Two Commercially Available Assays (Abbott RealTime HIV-1 and Roche Cobas AmpliPrep/Cobas TaqMan HIV-1 Version 2.0) To Quantify Low HIV-1 RNA Levels (1000 Copies/Milliliter): Comparison with Clinical Samples and NIBSC Working Reagent for Nucleic Acid Testing Assays

机译:两种商业化检测方法(Abbott RealTime HIV-1和Roche Cobas AmpliPrep / Cobas TaqMan HIV-1 2.0版)对低HIV-1 RNA水平(1000份/毫升)进行定量的能力:与临床样品和NIBSC工作试剂的比较核酸检测方法

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Concordance between molecular assays may be suboptimal at low HIV-1 viremia levels (<1,000 copies/ml); therefore, it may be difficult to define and compare virologic endpoints for successful and failed therapy. We compared two commercial assays (the Abbott RealTime HIV-1 and the Roche Cobas AmpliPrep/TaqMan HIV-1 version 2.0) for their ability to detect and quantify low viral loads. A comparison was performed using 167 residual clinical samples (with values ranging from “not detected” to 1,000 copies/ml, as measured by the Abbott assay) and the National Institute and Biological Standards and Control (NIBSC) HIV-1 RNA working reagent 1 for nucleic acid amplification techniques (NAT) assays (serially diluted to a range from 1 to 1,000 copies/ml). Quantitative results were compared using Lin's concordance correlation coefficient and a Bland-Altman plot. Concordance with the qualitative results was measured by Cohen's kappa statistic. With clinical samples, the degree of interassay concordance of the qualitative results at a 40-copies/ml HIV-1 RNA threshold was substantial (κ = 0.762); the correlation among the quantified samples was suboptimal (concordance correlation coefficient, 0.728; P < 0.0001); the mean difference of the values between the Roche and Abbott assays was 0.193 log10 copies/ml. Using the HIV-1 RNA working reagent 1 for NAT assays, the results provided by the Roche assay were, on average, 3 times higher than expected, while the Abbott assay showed high accuracy. The Roche assay was highly sensitive, being able to detect a level as low as 3.5 copies/ml HIV-1 RNA with 95% probability. The performance characteristics of each molecular assay should be taken into account when HIV-1 RNA threshold values for “virologic suppression,” “virologic failure,” “persistent low viral loads,” etc., are defined and indicated in the support of clinical decisions.
机译:在低HIV-1病毒血症水平(<1,000拷贝/ ml)下,分子分析之间的一致性可能不理想。因此,可能难以定义和比较成功和失败治疗的病毒学终点。我们比较了两种商业检测方法(Abbott RealTime HIV-1和Roche Cobas AmpliPrep / TaqMan HIV-1 2.0版)检测和定量低病毒量的能力。使用167个残留临床样品(通过Abbott分析测得的值,范围从“未检出”至1,000拷贝/ ml)和美国国家生物医学标准与对照(NIBSC)HIV-1 RNA工作试剂1进行了比较。用于核酸扩增技术(NAT)分析(血清稀释至1至1,000拷贝/ ml的范围)。使用Lin的一致性相关系数和Bland-Altman图比较定量结果。通过Cohen的kappa统计量来测量与定性结果的一致性。对于临床样品,在40拷贝/ ml HIV-1 RNA阈值下,定性结果的批间一致性程度很高(κ= 0.762);量化样本之间的相关性不佳(一致性相关系数为0.728; P <0.0001); Roche和Abbott分析之间值的平均差为0.193 log10拷贝/ ml。使用HIV-1 RNA工作试剂1进行NAT检测,罗氏检测提供的结果平均比预期高3倍,而雅培检测则显示出很高的准确性。罗氏分析法具有很高的灵敏度,能够以95%的概率检测到低至3.5拷贝/ ml HIV-1 RNA的水平。当定义了“ HIV-1 RNA”阈值用于“病毒抑制”,“病毒衰竭”,“持续的低病毒载量”等,并在临床决策支持中表明时,应考虑每种分子测定的性能特征。