Volumetric breast density measurement for personalized screening: accuracy reproducibility consistency and agreement with visual assessment

机译:体积乳房密度测量用于个性化筛查:准确性可重复性一致性并与视觉评估相符

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Assessment of breast density at the point of mammographic examination could lead to optimized breast cancer screening pathways. The onsite breast density information may offer guidance of when to recommend supplemental imaging for women in a screening program. A software application (Insight BD, Siemens Healthcare GmbH) for fast onsite quantification of volumetric breast density is evaluated. The accuracy of the method is assessed using breast tissue equivalent phantom experiments resulting in a mean absolute error of 3.84%. Reproducibility of measurement results is analyzed using 8427 exams in total, comparing for each exam (if available) the densities determined from left and right views, from cranio-caudal and medio-lateral oblique views, from full-field digital mammograms (FFDM) and digital breast tomosynthesis (DBT) data and from two subsequent exams of the same breast. Pearson correlation coefficients of 0.937, 0.926, 0.950, and 0.995 are obtained. Consistency of the results is demonstrated by evaluating the dependency of the breast density on women’s age. Furthermore, the agreement between breast density categories computed by the software with those determined visually by 32 radiologists is shown by an overall percentage agreement of 69.5% for FFDM and by 64.6% for DBT data. These results demonstrate that the software delivers accurate, reproducible, and consistent measurements that agree well with the visual assessment of breast density by radiologists.
机译:在乳房X光检查时评估乳房密度可导致优化的乳腺癌筛查途径。现场乳房密度信息可为筛查程序中何时建议女性补充成像提供指导。评估了用于快速现场定量乳房体积密度的软件应用程序(Insight BD,西门子医疗有限公司)。使用乳房组织等效体模实验评估该方法的准确性,得出的平均绝对误差为3.84%。总共使用了8427次检查来分析测量结果的可重复性,并比较每项检查(如果有)的密度,这些密度是从左,右视图,颅底和中外侧斜视图,全场数字乳房X线照片(FFDM)和数字乳房断层合成(DBT)数据以及同一乳房的两次后续检查。获得的皮尔逊相关系数为0.937、0.926、0.950和0.995。通过评估乳房密度对女性年龄的依赖性,证明了结果的一致性。此外,通过软件计算出的乳房密度类别与由32位放射科医生目测确定的密度之间的一致性,对于FFDM和DBT数据而言,总体百分比一致性为69.5%和64.6%。这些结果表明,该软件可提供准确,可重复且一致的测量结果,与放射科医生对乳房密度的视觉评估非常吻合。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号