Stretch calculated from grip distance accurately approximates mid-specimen stretch in large elastic arteries in uniaxial tensile tests

机译:根据抓地距离计算出的拉伸力在单轴拉伸试验中可以准确地近似大型弹性动脉的中试拉伸力

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The mechanical properties of vascular tissues affect hemodynamics and can alter disease progression. The uniaxial tensile test is a simple and effective method for determining the stress-strain relationship in arterial tissue ex vivo. To enable calculation of strain, stretch can be measured directly with image tracking of markers on the tissue or indirectly from the distance between the grips used to hold the specimen. While the imaging technique is generally considered more accurate, it also requires more analysis, and the grip distance method is more widely used. The purpose of this study is to compare the stretch of the testing specimen calculated from the grip distance method to that obtained from the imaging method for canine descending aortas and large proximal pulmonary arteries. Our results showed a significant difference in stretch between the two methods; however, this difference was consistently less than 2%. Therefore, the grip distance method is an accurate approximation of the stretch in large elastic arteries in the uniaxial tensile test.
机译:血管组织的机械特性会影响血液动力学,并可能改变疾病进展。单轴拉伸试验是一种简单有效的方法,可用于体外测定动脉组织中的应力-应变关系。为了能够计算应变,可以直接在组织上标记的图像跟踪中测量拉伸,也可以从用于固定样品的夹具之间的距离间接测量拉伸。虽然通常认为成像技术更准确,但它也需要更多分析,并且抓地距离方法得到更广泛的应用。这项研究的目的是比较从抓地距离方法计算得出的测试样本的拉伸与从成像方法获得的犬降主动脉和大型近端肺动脉的拉伸的比较。我们的结果表明两种方法之间的拉伸差异显着。但是,这一差异始终小于2%。因此,握持距离法是单轴拉伸试验中大弹性动脉中拉伸的精确近似值。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号