An improvement in IMRT QA results and beam matching in linacs using statistical process control

机译:使用统计过程控制改进直线加速器的IMRT QA结果和光束匹配

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study is to apply the principles of statistical process control (SPC) in the context of patient specific intensity‐modulated radiation therapy (IMRT) QA to set clinic‐specific action limits and evaluate the impact of changes to the multileaf collimator (MLC) calibrations on IMRT QA results. Ten months of IMRT QA data with 247 patient QAs collected on three beam‐matched linacs were retrospectively analyzed with a focus on the gamma pass rate (GPR) and the average ratio between the measured and planned doses. Initial control charts and action limits were calculated. Based on this data, changes were made to the leaf gap parameter for the MLCs to improve the consistency between linacs. This leaf gap parameter is tested monthly using a MLC sweep test. A follow‐up dataset with 424 unique QAs were used to evaluate the impact of the leaf gap parameter change. The initial data average GPR was 98.6% with an SPC action limit of 93.7%. The average ratio of doses was 1.003, with an upper action limit of 1.017 and a lower action limit of 0.989. The sweep test results for the linacs were 1.8%,0%, and +1.2% from nominal. After the adjustment of the leaf gap parameter, all sweep test results were within 0.4% of nominal. Subsequently, the average GPR was 99.4% with an SPC action limit of 97.3%. The average ratio of doses was 0.997 with an upper action limit of 1.011 and a lower action limit of 0.981. Applying the principles of SPC to IMRT QA allowed small differences between closely matched linacs to be identified and reduced. Ongoing analysis will monitor the process and be used to refine the clinical action limits for IMRT QA.PACS number: 87.55.Qr
机译:这项研究的目的是在患者特定的强度调制放射治疗(IMRT)QA的背景下应用统计过程控制(SPC)的原理来设置临床特定的作用极限并评估变化对多叶准直仪的影响( IMRT QA结果的MLC)校准。回顾性分析了在三个光束匹配的直线加速器上收集的247个患者QA的10个月的IMRT QA数据,重点是伽玛通过率(GPR)以及测量剂量与计划剂量之间的平均比率。计算初始控制图和动作极限。基于此数据,对MLC的叶间隙参数进行了更改,以提高直线加速器之间的一致性。每月使用MLC扫描测试对该叶间距参数进行测试。使用具有424个唯一QA的后续数据集来评估叶间隙参数变化的影响。初始数据平均GPR为98.6%,SPC操作限制为93.7%。平均剂量比为1.003,上限为1.017,下限为0.989。直线加速器的扫描测试结果为 - 1.8 0 + 1.2 (标称值)。调整叶片间隙参数后,所有扫掠测试结果均在标称值的0.4%以内。随后,平均GPR为99.4%,SPC操作限制为97.3%。平均剂量比为0.997,上限为1.011,下限为0.981。将SPC原理应用于IMRT QA可以识别和减少紧密匹配的直线加速器之间的微小差异。正在进行的分析将监视该过程,并用于完善IMRT QA.PACS编号:87.55.Qr的临床行动限制

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号