A software tool to automatically assure and report daily treatment deliveries by a cobalt‐60 radiation therapy device

机译:一种软件工具可自动确保和报告钴60放射治疗仪的每日治疗量

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aims of this study were to develop a method for automatic and immediate verification of treatment delivery after each treatment fraction in order to detect and correct errors, and to develop a comprehensive daily report which includes delivery verification results, daily image‐guided radiation therapy (IGRT) review, and information for weekly physics reviews. After systematically analyzing the requirements for treatment delivery verification and understanding the available information from a commercial MRI‐guided radiotherapy treatment machine, we designed a procedure to use 1) treatment plan files, 2) delivery log files, and 3) beam output information to verify the accuracy and completeness of each daily treatment delivery. The procedure verifies the correctness of delivered treatment plan parameters including beams, beam segments and, for each segment, the beam‐on time and MLC leaf positions. For each beam, composite primary fluence maps are calculated from the MLC leaf positions and segment beam‐on time. Error statistics are calculated on the fluence difference maps between the plan and the delivery. A daily treatment delivery report is designed to include all required information for IGRT and weekly physics reviews including the plan and treatment fraction information, daily beam output information, and the treatment delivery verification results. A computer program was developed to implement the proposed procedure of the automatic delivery verification and daily report generation for an MRI guided radiation therapy system. The program was clinically commissioned. Sensitivity was measured with simulated errors. The final version has been integrated into the commercial version of the treatment delivery system. The method automatically verifies the EBRT treatment deliveries and generates the daily treatment reports. Already in clinical use for over one year, it is useful to facilitate delivery error detection, and to expedite physician daily IGRT review and physicist weekly chart review.PACS number(s): 87.55.km
机译:这项研究的目的是开发一种方法,用于在每个治疗阶段之后自动并立即验证治疗的执行情况,以发现并纠正错误;并制定一份全面的每日报告,其中包括交付验证结果,每日图像引导的放射治疗( IGRT)评论,以及每周物理评论的信息。在系统地分析了治疗交付验证的要求并了解了商用MRI引导的放射治疗设备的可用信息之后,我们设计了使用1)治疗计划文件,2)交付日志文件和3)束输出信息进行验证的过程每天日常治疗的准确性和完整性。该程序将验证所提供的治疗计划参数的正确性,包括光束,光束分段以及每个分段的光束开启时间和MLC叶位置。对于每个光束,根据MLC叶片位置和分段光束接通时间计算出合成的主注量图。错误统计信息是根据计划和交付之间的注量差异图计算的。每日治疗提供报告旨在包含IGRT所需的所有信息,以及每周物理检查,包括计划和治疗分数信息,每日束输出信息以及治疗提供验证结果。开发了一个计算机程序来实施MRI指导的放射治疗系统的自动交付验证和每日报告生成的建议程序。该程序已临床调试。用模拟误差测量灵敏度。最终版本已集成到治疗交付系统的商业版本中。该方法会自动验证EBRT的治疗效果并生成每日治疗报告。在临床上已经使用了一年多,它对促进传递错误检测,加快医师的每日IGRT复查和物理学家的每周图表复查非常有用。PACS编号:87.55.km

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号