Measurement of wire deflection on loading may indicate union in Ilizarov constructs an in vitro model

机译:测量导线在载荷下的挠度可能表明在Ilizarov构造中的结合这是一种体外模型

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

No entirely reliable method exists for assessing union during Ilizarov treatment. Premature removal results in potential treatment failure; hence, alternative methods warrant investigation. Wire deflection might provide an indication of fracture site deformation on weight bearing, indicating progress towards union. This study aimed to test a method for assessing wire deflection within an Ilizarov frame. (1) To assess the repeatability of our novel measurement method in measuring wire deflection within an Ilizarov frame in vitro. (2) To compare the amount of wire deflection in an unstable model with that in an intact bone model. (3) To assess accuracy of this method by comparing wire deflection measured with overall machine extension. Tests were performed on clinical grade-tensioned fine wire 4-ring Ilizarov constructs stabilising a simulated fracture, with and without an unstable defect. Models were sequentially loaded to 700 N using an Instron testing machine. A digital depth gauge attached to the superior ring measured relative wire displacement at the ring closest to the fracture. Tests were repeated 3 times. (1) Both unstable and stable bone models produced highly repeatable load deformation curves (R2 = 0.98 and 0.99). (2) In the unstable model, wires tensioned at 882 and 1274 N produced mean maximum deflections of 2.41 and 2.69 mm compared with 0.05 and 0.04 mm in the intact bone model (significant p < 0.0001). (3) Wire deflection and machine extension results were strongly correlated (r = 0.99). A measurable difference in wire deflection between stable and unstable situations exists using this method which appears accurate and repeatable, with clear correlation between displacement and load and displacement and machine extension. This approach might be clinically applicable, and further clinical testing is required.
机译:在Ilizarov治疗期间,尚无完全可靠的方法来评估结合。过早清除会导致潜在的治疗失败;因此,其他方法值得调查。钢丝变形可能会显示承重上的断裂部位变形,从而表明朝着结合的方向发展。这项研究旨在测试一种用于评估Ilizarov框架内钢丝变形的方法。 (1)评估我们新颖的测量方法在体外测量Ilizarov框架内的导线挠度的可重复性。 (2)比较不稳定模型和完整骨模型中的线偏斜量。 (3)通过比较测得的导线挠度与整个机器延伸范围来评估该方法的准确性。对临床等级张紧的细线4环Ilizarov构造物进行了测试,该构造物可稳定有或没有不稳定缺陷的模拟骨折。使用Instron测试机将模型顺序加载到700 N.连接到上环的数字深度计可测量最靠近裂缝的环上的相对线位移。重复测试3次。 (1)不稳定和稳定的骨骼模型均产生高度可重复的载荷变形曲线(R 2 = 0.98和0.99)。 (2)在不稳定模型中,在882和1274 N处张紧的钢丝产生的平均最大挠度为2.41和2.69 mm,而在完整的骨骼模型中则为0.05和0.04 mm(p显着小于0.0001)。 (3)导线挠度和机器延伸结果密切相关(r = 0.99)。使用这种方法,在稳定状态和不稳定状态之间存在可测量的导线挠度差异,该差异看起来准确且可重复,并且位移和负载以及位移和机器延伸之间具有明显的相关性。这种方法可能在临床上适用,并且需要进一步的临床测试。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号