Thousands of operations are annually guided with computer assisted surgery (CAS) technologies. As the use of these devices is rapidly increasing, the reliability of the devices becomes ever more critical. The problem of accuracy assessment of the devices has thus become relevant. During the past five years, over 200 hazardous situations have been documented in the MAUDE database during operations using these devices in the field of neurosurgery alone. Had the accuracy of these devices been periodically assessed pre-operatively, many of them might have been prevented.The technical accuracy of a commercial navigator enabling the use of both optical (OTS) and electromagnetic (EMTS) tracking systems was assessed in the hospital setting using accuracy assessment tools and methods developed by the authors of this paper. The technical accuracy was obtained by comparing the positions of the navigated tool tip with the phantom accuracy assessment points. Each assessment contained a total of 51 points and a region of surgical interest (ROSI) volume of 120x120x100 mm roughly mimicking the size of the human head.The error analysis provided a comprehensive understanding of the trend of accuracy of the surgical navigator modalities. This study showed that the technical accuracies of OTS and EMTS over the pre-determined ROSI were nearly equal. However, the placement of the particular modality hardware needs to be optimized for the surgical procedure. New applications of EMTS, which does not require rigid immobilization of the surgical area, are suggested.
机译:每年都有数以千计的手术在计算机辅助手术(CAS)技术的指导下进行。随着这些设备的使用迅速增加,设备的可靠性变得越来越重要。因此,设备的准确性评估问题变得很重要。在过去的五年中,仅在神经外科领域使用这些设备进行操作时,MAUDE数据库中记录了200多种危险情况。如果在术前定期评估这些设备的准确性,可能会阻止许多此类行为。在医院环境中评估了同时使用光学(OTS)和电磁(EMTS)跟踪系统的商用导航仪的技术准确性。使用本文作者开发的准确性评估工具和方法。通过将导航的工具提示的位置与幻像精度评估点进行比较来获得技术精度。每次评估共计51个要点,手术区域(ROSI)的大小为120x120x100 mm,大致模拟人的头部大小。误差分析提供了对手术导航仪模式准确性趋势的全面了解。这项研究表明,OTS和EMTS在预定ROSI上的技术精度几乎相等。但是,需要针对手术过程优化特定模态硬件的放置。建议使用EMTS的新应用,该技术不需要牢固固定手术区域。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号