PurposeA prototype QC phantom system and analysis process were developed to characterize the spectral capabilities of a fast kV-switching dual-energy computed tomography (DECT) scanner. This work addresses the current lack of quantitative oversight for this technology, with the goal of identifying relevant scan parameters and test metrics instrumental to the development of a dual-energy quality control (DEQC).
机译:目的开发了原型QC幻象系统和分析过程,以表征快速kV切换双能计算机断层扫描(DECT)扫描仪的光谱功能。这项工作解决了目前对该技术缺乏量化监督的问题,目的是确定有助于开发双能质量控制(DEQC)的相关扫描参数和测试指标。