首页> 美国卫生研究院文献>Medical Physics >A fast angle-dependent analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems
【2h】

A fast angle-dependent analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems

机译:快速与角度相关的CsI探测器响应分析模型用于优化3D X射线乳腺成像系统

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Purpose: Accurate models of detector blur are crucial for performing meaningful optimizations of three-dimensional (3D) x-ray breast imaging systems as well as for developing reconstruction algorithms that faithfully reproduce the imaged object anatomy. So far, x-ray detector blur has either been ignored or modeled as a shift-invariant symmetric function for these applications. The recent development of a Monte Carlo simulation package called MANTIS has allowed detailed modeling of these detector blur functions and demonstrated the magnitude of the anisotropy for both tomosynthesis and breast CT imaging systems. Despite the detailed results that MANTIS produces, the long simulation times required make inclusion of these results impractical in rigorous optimization and reconstruction algorithms. As a result, there is a need for detector blur models that can be rapidly generated.>Methods: In this study, the authors have derived an analytical model for deterministic detector blur functions, referred to here as point response functions (PRFs), of columnar CsI phosphor screens. The analytical model is x-ray energy and incidence angle dependent and draws on results from MANTIS to indirectly include complicated interactions that are not explicitly included in the mathematical model. Once the mathematical expression is derived, values of the coefficients are determined by a two-dimensional (2D) fit to MANTIS-generated results based on a figure-of-merit (FOM) that measures the normalized differences between the MANTIS and analytical model results averaged over a region of interest. A smaller FOM indicates a better fit. This analysis was performed for a monochromatic x-ray energy of 25 keV, a CsI scintillator thickness of 150 μm, and four incidence angles (0°, 15°, 30°, and 45°).>Results: The FOMs comparing the analytical model to MANTIS for these parameters were 0.1951±0.0011, 0.1915±0.0014, 0.2266±0.0021, and 0.2416±0.0074 for 0°, 15°, 30°, and 45°, respectively. As a comparison, the same FOMs comparing MANTIS to 2D symmetric Gaussian fits to the zero-angle PRF were 0.6234±0.0020, 0.9058±0.0029, 1.491±0.012, and 2.757±0.039 for the same set of incidence angles. Therefore, the analytical model matches MANTIS results much better than a 2D symmetric Gaussian function. A comparison was also made against experimental data for a 170 μm thick CsI screen and an x-ray energy of 25.6 keV. The corresponding FOMs were 0.3457±0.0036, 0.3281±0.0057, 0.3422±0.0023, and 0.3677±0.0041 for 0°, 15°, 30°, and 45°, respectively. In a previous study, FOMs comparing the same experimental data to MANTIS PRFs were found to be 0.2944±0.0027, 0.2387±0.0039, 0.2816±0.0025, and 0.2665±0.0032 for the same set of incidence angles.>Conclusions: The two sets of derived FOMs, comparing MANTIS-generated PRFs and experimental data to the analytical model, demonstrate that the analytical model is able to reproduce experimental data with a FOM of less than two times that comparing MANTIS and experimental data. This performance is achieved in less than one millionth the computation time required to generate a comparable PRF with MANTIS. Such small computation times will allow for the inclusion of detailed detector physics in rigorous optimization and reconstruction algorithms for 3D x-ray breast imaging systems.
机译:>目的:准确的探测器模糊模型对于执行三维(3D)X射线乳房成像系统的有意义的优化以及开发忠实再现成像对象解剖结构的重建算法至关重要。到目前为止,对于这些应用,x射线检测器的模糊要么被忽略,要么被建模为位移不变对称函数。名为MANTIS的蒙特卡罗仿真程序包的最新开发已经实现了对这些探测器模糊功能的详细建模,并证明了断层合成和乳腺CT成像系统的各向异性大小。尽管MANTIS产生了详细的结果,但所需的较长的仿真时间使得将这些结果包含在严格的优化和重建算法中是不切实际的。结果,需要可以快速生成的检测器模糊模型。>方法:在这项研究中,作者得出了确定性检测器模糊函数的分析模型,在此称为点响应。柱状CsI磷光屏的功能(PRF)。该分析模型与X射线能量和入射角有关,并利用MANTIS的结果间接包含数学模型中未明确包含的复杂相互作用。推导数学表达式后,系数的值由基于衡量因数(FOM)的MANTIS和分析模型结果之间的标准化差异的二维(2D)拟合到MANTIS生成的结果来确定在感兴趣的区域进行平均。较小的FOM表示更合适。此分析是针对25keV的单色X射线能量,150μm的CsI闪烁体厚度和四个入射角(0°,15°,30°和45°)进行的。>结果:对于这些参数,将分析模型与MANTIS进行比较的FOM分别为0°,15°,30°和45°,分别为0.1951±0.0011、0.1915±0.0014、0.2266±0.0021和0.2416±0.0074。作为比较,对于同一组入射角,将MANTIS与2D对称高斯拟合与零角PRF进行比较的相同FOM为0.6234±0.0020、0.9058±0.0029、1.491±0.012和2.757±0.039。因此,与MANTIS结果匹配的解析模型要比2D对称高斯函数好得多。还针对170μm厚的CsI屏幕和25.6 keV的X射线能量与实验数据进行了比较。对于0°,15°,30°和45°,相应的FOM分别为0.3457±0.0036、0.3281±0.0057、0.3422±0.0023和0.3677±0.0041。在先前的研究中,对于同一组入射角,将相同实验数据与MANTIS PRF进行比较的FOMs为0.2944±0.0027、0.2387±0.0039、0.2816±0.0025和0.2665±0.0032。>结论:两组衍生的FOM将MANTIS生成的PRF和实验数据与分析模型进行了比较,证明了该分析模型能够再现FOM小于比较MANTIS和实验数据的两倍的实验数据。在用MANTIS生成可比的PRF所需的计算时间不到百万分之一的情况下就可以实现这种性能。如此小的计算时间将使详细的检测器物理特性包含在针对3D X射线乳房成像系统的严格优化和重建算法中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号