Validation of an in-vivo proton beam range check method in an anthropomorphic pelvic phantom using dose measurements

机译:使用剂量测量在拟人化骨盆模型中体内质子束范围检查方法的验证

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose:In-vivo dosimetry and beam range verification in proton therapy could play significant role in proton treatment validation and improvements. In-vivo beam range verification, in particular, could enable new treatment techniques one of which could be the use of anterior fields for prostate treatment instead of opposed lateral fields as in current practice. This paper reports validation study of an in-vivo range verification method which can reduce the range uncertainty to submillimeter levels and potentially allow for in-vivo dosimetry.
机译:目的:质子治疗中的体内剂量测定和射束范围验证可在质子治疗的验证和改进中发挥重要作用。尤其是,体内射束范围验证可以启用新的治疗技术,其中之一可以是使用前场进行前列腺治疗,而不是像目前的实践中那样使用相对的侧场。本文报告了一种体内范围验证方法的验证研究,该方法可以将范围不确定性降低到亚毫米水平,并有可能进行体内剂量测定。