Saturation curves for free-air ionization chambers with electrode gap widths of 4.2, 8.4 and 18 mm were obtained for 10 and 15 keV undulator synchrotron radiation thinned with a 230 Hz rotating-disk chopper. Ion recombination in free-air ionization chambers was found to be inversely proportional to the applied electric field, and an expression that satisfactorily reproduced the ion-recombination rate is determined. A comparison of the expressions for continuous and pulsed X-rays revealed that chopped high-intensity X-rays require a higher voltage to attain saturation when the product of the pulse width and electric field exceeds a value that depends on the X-ray energy. This behaviour was observed explicitly for 10 keV X-rays in measurements with the ionization chamber placed before and after the chopper.
机译:对于用230 Hz转盘斩波器变薄的10 and和15 keV波动加速器同步辐射,获得了电极间隙宽度为4.2、8.4和18 mm的自由空气电离室的饱和曲线。发现在自由空气电离室中的离子重组与所施加的电场成反比,并且确定了令人满意地再现离子重组速率的表达式。比较连续X射线和脉冲X射线的表达式,可以发现,当脉冲宽度和电场的乘积超过取决于X射线能量的值时,斩波的高强度X射线需要更高的电压才能达到饱和。对于10 keV X射线,在将电离室放置在斩波器之前和之后的测量中,明确观察到了这种行为。