首页> 美国卫生研究院文献>Journal of Synchrotron Radiation >A small and robust active beamstop for scattering experiments on high-brilliance undulator beamlines
【2h】

A small and robust active beamstop for scattering experiments on high-brilliance undulator beamlines

机译:小型坚固的主动式光束停止器用于高亮度波状起伏器光束线上的散射实验

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A small active in-vacuum beamstop has been developed to monitor the flux of intense third-generation synchrotron X-ray beams protecting the downstream detector from the direct beam. Standard active beamstops, where a built-in diode directly absorbs the beam, have limitations in size and lifetime. In the present design, a silicon PIN diode detects the photons back-scattered from a cavity in the beamstop. This approach drastically reduces the radiation dose on the diode and thus increases its lifetime. The beamstop with a diameter of 2 mm has been fabricated to meet the requirements for the P12 bioSAXS beamline of EMBL Hamburg at PETRA III (DESY). The beamstop is in regular user operation at the beamline and displays a good response over the range of energies tested (6–20 keV). Further miniaturization of the diode is easily possible as its size is not limited by the PIN diode used.
机译:已经开发出一种小型有源真空束停止器,以监视强烈的第三代同步加速器X射线束的通量,从而保护下游探测器免受直射束的侵害。内置二极管直接吸收光束的标准有源光束停止器在尺寸和使用寿命方面都有局限性。在本设计中,硅PIN二极管检测从光束停止器中的腔反向散射的光子。这种方法大大减少了二极管上的辐射剂量,从而延长了其寿命。直径为2 mm的光束止挡件的制造符合PETRA III(DESY)上EMBL Hamburg的P12 bioSAXS光束线的要求。光束停止器在光束线上的常规用户操作中,在测试的能量范围(6–20 keV)上显示出良好的响应。由于二极管的尺寸不受所用PIN二极管的限制,因此很容易实现二极管的进一步小型化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号