首页> 美国卫生研究院文献>Acta Orthopaedica >CD3+ lymphocytosis in the peri-implant membrane of 222 loosened joint endoprostheses depends on the tribological pairing
【2h】

CD3+ lymphocytosis in the peri-implant membrane of 222 loosened joint endoprostheses depends on the tribological pairing

机译:222个松动的关节内假体的植入物周围膜中的CD3 +淋巴细胞增多取决于摩擦学配对

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Background and purpose — The most frequent cause of arthroplasty failure is aseptic loosening—often induced by particles. Abrasion material triggers inflammatory reactions with lymphocytic infiltration and the formation of synovial-like interface membranes (SLIM) in the bone–implant interface. We analyzed CD3 quantities in SLIM depending on articulating materials and possible influences of proven material allergies on CD3 quantities.>Patients and methods — 222 SLIM probes were obtained from revision surgeries of loosened hip and knee arthroplasties. SLIM cases were categorized according to the SLIM-consensus classification and to the particle algorithm. The CD3 quantities were analyzed immunohistochemically, quantified, and correlated to the particle types.>Results — Metal–metal pairings showed the highest CD3 quantities (mean 1,367 counted cells). CD3 quantities of metal–polyethylene (mean 243), ceramic–polyethylene (mean 182), and ceramic–ceramic pairings (mean 124) were significantly smaller. Patients with contact allergy to implant materials had high but not statistically significantly higher CD3 quantities than patients without allergies. For objective assessment of the CD3 response as result of a pronounced inflammatory reaction with high lymphocytosis (adverse reaction), a defined CD3 quantity per high power field was established, the “CD3 focus score” (447 cells/0.3 mm2, sensitivity 0.92; specificity 0.90; positive predictive value 0.71; negative predictive value 0.98).>Interpretation — The high CD3 quantities for metal–metal pairings may be interpreted as substrate for previously described adverse reactions that cause severe peri-implant tissue destruction and SLIM formation. It remains unclear whether the low CD3 quantities with only slight differences in the various non-metal–metal pairings and documented contact allergies to implant materials have a direct pathogenetic relevance.
机译:>背景和目的-人工关节置换失败的最常见原因是无菌性松动,通常是由颗粒引起的。磨损材料会触发炎症反应,并伴有淋巴细胞浸润,并在骨-植入物界面形成滑膜样界面膜(SLIM)。我们分析了SLIM中CD3的数量,具体取决于关节材料以及已证明的材料过敏对CD3数量的可能影响。>患者和方法--222根SLIM探针是从松动的髋和膝关节置换术的翻修手术中获得的。根据SLIM共识分类和粒子算法对SLIM案例进行了分类。对CD3的数量进行了免疫组织化学分析,定量并与颗粒类型相关。>结果-金属-金属配对显示出最高的CD3数量(平均1,367个计数细胞)。金属-聚乙烯(平均243),陶瓷-聚乙烯(平均182)和陶瓷-陶瓷配对(平均124)的CD3数量要少得多。对植入材料有接触过敏的患者比没有过敏的患者有较高的CD3量,但统计学上却不显着更高。为了客观评估由于明显的炎症反应和高淋巴细胞增多(不良反应)而导致的CD3反应,确定了每个高倍视野的CD3量,即“ CD3聚焦评分”(447个细胞/0.3?mm 2 < / sup>,灵敏度0.92;特异性0.90;阳性预测值0.71;阴性预测值0.98)。>解释-金属-金属配对的高CD3量可解释为先前所述不良反应的底物导致严重的种植体周围组织破坏和SLIM形成。尚不清楚低CD3量是否在各种非金属-金属配对中仅存在微小差异,并且有记载的对植入材料的接触过敏是否具有直接的病原学意义。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号