The isolated circuit diathermy.

机译:隔离电路透热。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

When using the conventional diathermy generator in surgery failure to apply the plate electrode can always present a serious risk of a thermoelectrical burn at any point where the patient makes contact with an earthed object on the operating table. The recent introduction of the earth-free (isolated circuit) diathermy should, under most conditions, provide complete protection against such risks. However, there are still certain circumstances under which even the isolated circuit may give rise to a thermoelectrical burn. Two such hypothetical examples are described; the sequences of events for these are fairly frequent occurrence.
机译:当在手术中使用常规的透热发生器时,在患者与手术台上的接地物体接触的任何位置,施加平板电极总是会存在严重的热电灼伤危险。在大多数情况下,最近引入的无接地(隔离电路)透热疗法应能为此类风险提供全面的保护。但是,在某些情况下,即使隔离的电路也可能引起热电灼伤。描述了两个这样的假设示例;这些事件的发生顺序相当频繁。