Comparison between a New Optical Biometry Device and an Anterior Segment Optical Coherence Tomographer for Measuring Central Corneal Thickness and Anterior Chamber Depth

机译:新型光学生物测定仪与前节光学相干断层扫描仪测量中央角膜厚度和前房深度的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose. To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD). Methods. Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using paired t-tests, Bland-Altman plots, and 95% limits of agreement (LoA). Results. The mean CCT, ACD, and AD measured by AL-Scan were 538.59 ± 27.37 μm, 3.70 ± 0.30 mm, and 3.16 ± 0.30 mm, respectively. The mean values obtained by the Visante OCT were 536.14 ± 26.61 μm for CCT, 3.71 ± 0.29 mm for ACD, and 3.17 ± 0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference = 2.45 ± 6.07 μm, P < 0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively. Conclusions. Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.
机译:目的。为了比较新型光学生物计(AL-Scan,日本爱知县Nidek公司)与前节光学相干断层扫描仪(Visante AS-OCT,卡尔·蔡司·梅迪泰克,美国都柏林)来测量中央角膜厚度(CCT),前房深度(ACD)和房水深度(AD)。方法。在这项前瞻性研究中,对63名正常受试者的63只眼睛进行了AL-Scan和Visante AS-OCT检查。使用这两种设备对每个受试者的一只眼睛进行三次测量,以记录其CCT,ACD和AD。所有程序均由同一操作员执行。使用配对t检验,Bland-Altman图和95%的协议限制(LoA)评估了两种设备之间的协议。结果。通过AL-Scan测得的平均CCT,ACD和AD分别为538.59±27.37μm,3.70±0.30 mm和3.16±0.30 mm。通过Visante OCT获得的平均值对于CCT是536.14±26.61μm,对于ACD是3.71±0.29 mm,对于AD是3.17±0.29 mm。 AL-Scan的平均CCT高于Visante AS-OCT的平均CCT(差异= 2.45±6.07μm,P <0.05)。 ACD和AD测量值的差异无统计学意义。 CCT,ACD和AD的95%LoA分别在-9.44和14.35μm之间,-0.15和0.12μmm之间以及-0.15和0.12μmm之间。结论。由于这两种设备在测量CCT,ACD和AD方面具有可比性,因此它们的结果可在临床中互换使用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号