Hollow optical-fiber based infrared spectroscopy for measurement of blood glucose level by using multi-reflection prism

机译:基于中空光纤的红外光谱仪通过多反射棱镜测量血糖水平

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A mid-infrared attenuated total reflection (ATR) spectroscopy system employing hollow optical fibers and a trapezoidal multi-reflection ATR prism has been developed to measure blood glucose levels. Using a multi-reflection prism brought about higher sensitivity, and the flat and wide contact surface of the prism resulted in higher measurement reproducibility. An analysis of in vivo measurements of human inner lip mucosa revealed clear signatures of glucose in the difference spectra between ones taken during the fasting state and ones taken after ingestion of glucose solutions. A calibration plot based on the absorption peak at 1155 cm−1 that originates from the pyranose ring structure of glucose gave measurement errors less than 20%.
机译:已经开发出一种采用中空光纤和梯形多反射ATR棱镜的中红外衰减全反射(ATR)光谱系统,用于测量血糖水平。使用多反射棱镜带来更高的灵敏度,并且棱镜的平坦且宽阔的接触表面带来了更高的测量重现性。对人内唇粘膜的体内测量结果分析显示,在禁食状态下摄取的葡萄糖与摄入葡萄糖溶液后摄取的葡萄糖之间的差异光谱中,葡萄糖具有清晰的特征。基于源自葡萄糖的吡喃糖环结构的在1155 cm -1 处的吸收峰的校准图给出的测量误差小于20%。