Iris clipping of a diode laser beam when performing retinal photocoagulation.

机译:进行视网膜光凝术时二极管激光束的虹膜削波。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Microlase infrared diode laser has recently become available for retinal photocoagulation. It uses a laser beam with a large convergence angle (23 degrees). This gives rise to a laser beam with a relatively large cross section at the subject's pupil. In this study the pupillary area required to view the fundus at the slit-lamp with a 90-dioptre lens is determined. It is found that the pupillary area required to view the fundus is smaller than the cross section of the Microlase laser beam at the pupil. Thus iris clipping of the laser beam can occur even though the surgeon has a clear and bright view of the fundus. Care should be taken to use the Microlase only in the presence of a widely dilated pupil.
机译:最近,Microlase红外二极管激光器可用于视网膜光凝。它使用会聚角大(23度)的激光束。这在对象的瞳孔处产生具有相对较大横截面的激光束。在这项研究中,确定了用90倍屈光度透镜在裂隙灯处观察眼底所需的瞳孔面积。发现观察眼底所需的瞳孔面积小于瞳孔处的Microlase激光束的横截面。因此,即使外科医生对眼底有清晰而明亮的视野,也可能发生激光束的虹膜削波。仅在瞳孔散大的情况下才应小心使用Microlase。