首页> 美国卫生研究院文献>Case Reports in Cardiology >A Rare Case of Complete Fragmentation of Pacemaker Lead after a High-Velocity Theme Park Ride
【2h】

A Rare Case of Complete Fragmentation of Pacemaker Lead after a High-Velocity Theme Park Ride

机译:在一次高速主题公园骑行之后出现起搏器铅完全碎裂的罕见情况

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Pacemaker lead fracture is one of the most common causes of pacemaker malfunction and is most frequently associated with weight lifting or chest trauma. These patients usually present with symptoms of dizziness, syncope, chest discomfort, and palpitations or less commonly with extracardiac symptoms. Diagnosis is made by ECG and careful review of chest imaging such as chest X-ray or fluoroscopy. Treatment involves placement of a new lead with or without extraction of the fractured lead. We present an interesting case of complete severance of the tip of a dual-chamber pacemaker's atrial lead after a high-velocity theme park ride. In our case, the fracture occurred during amusement park rides and went undiagnosed until the patient presented for routine pacemaker evaluation. This case indicates that extreme physical forces in the absence of direct trauma, such as during amusement park rides, may result in lead fractures and patients with pacemakers should be cautioned regarding such activities.
机译:起搏器铅断裂是起搏器故障的最常见原因之一,最常见于举重或胸部创伤。这些患者通常表现出头晕,晕厥,胸部不适和心的症状,或较少出现心外膜症状。通过心电图进行诊断,并仔细检查胸部影像,例如胸部X射线或荧光检查。治疗包括放置新的铅,有或没有取出断裂的铅。我们提出了一个有趣的案例,即在高速主题公园游乐设施之后,双室起搏器心房末端的末端被完全切断。在我们的案例中,骨折发生在游乐园游乐设施中,直到患者进行常规起搏器评估才被诊断。这种情况表明,在没有直接创伤的情况下(例如在游乐园中骑乘时)过大的体力可能会导致铅骨折,应提醒心脏起搏器患者注意此类活动。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号