Parallel imaging performance investigation of an 8-channel common-mode differential-mode (CMDM) planar array for 7T MRI

机译:用于7T MRI的8通道共模差模(CMDM)平面阵列的并行成像性能研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

An 8-channel planar phased array was proposed based on the common-mode differential-mode (CMDM) structure for ultrahigh field MRI. The parallel imaging performance of the 8-channel CMDM planar array was numerically investigated based on electromagnetic simulations and Cartesian sensitivity encoding (SENSE) reconstruction. The signal-to-noise ratio (SNR) of multichannel images combined using root-sum-of-squares (rSoS) and covariance weighted root-sum-of-squares (Cov-rSoS) at various reduction factors were compared between 8-channel CMDM array and 4-channel CM and DM array. The results of the study indicated the 8-channel CMDM array excelled the 4-channel CM and DM in SNR. The g-factor maps and artifact power were calculated to evaluate parallel imaging performance of the proposed 8-channel CMDM array. The artifact power of 8-channel CMDM array was reduced dramatically compared with the 4-channel CM and DM arrays demonstrating the parallel imaging feasibility of the CMDM array.
机译:提出了一种基于共模差模(CMDM)结构的8通道平面相控阵,用于超高场MRI。基于电磁仿真和笛卡尔灵敏度编码(SENSE)重建,对8通道CMDM平面阵列的并行成像性能进行了数值研究。比较了在8通道之间使用各种减少因子使用平方根均方根(rSoS)和协方差加权平方根均方根和(Cov-rSoS)组合的多通道图像的信噪比(SNR) CMDM阵列和4通道CM和DM阵列。研究结果表明,8通道CMDM阵列的信噪比优于4通道CM和DM。计算了g因子图和伪影功率,以评估建议的8通道CMDM阵列的并行成像性能。与4通道CM和DM阵列相比,8通道CMDM阵列的伪像功率大大降低,证明了CMDM阵列并行成像的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号